: Fracture healing is currently assessed through qualitative evaluation of radiographic images, which is highly subjective in nature. Radiographs can only provide snapshots in time, which are limited due to logistics and radiation exposure. We recently proposed assessing the bone healing status through continuous monitoring of the implant load, utilizing an implanted sensor system, the Fracture Monitor. The device telemetrically transmits statistically derived implant parameters via the patient's mobile phone to assist physicians in diagnostics and treatment decision-making. This preclinical study aims to systematically investigate the device safety and performance in an animal setting. : Mid-shaft tibial osteotomies of different sizes (0.6-30 mm) were created in eleven Swiss mountain sheep. The bones were stabilized with either a conventional Titanium or stainless-steel locking plate equipped with a Fracture Monitor. Data were continuously collected over the device's lifetime. Conventional radiographs and clinical CT scans were taken longitudinally over the study period. The radiographs were systematically scored and CTs were evaluated for normalized bone volume in the defect. The animals were euthanized after 9 months. The sensor output was correlated with the radiologic parameters. Tissue samples from the device location were histologically examined. : The sensors functioned autonomously for 6.5-8.4 months until energy depletion. No macroscopic or microscopic adverse effects from device implantation were observed. The relative implant loads at 4 and 8 weeks post-operation correlated significantly with the radiographic scores and with the normalized bone volume metric. : Continuous implant load monitoring appears as a relevant approach to support and objectify fracture healing assessments and carries a strong potential to enable patient-tailored rehabilitation in the future.

摘要

骨折愈合目前通过对射线图像的定性评估来进行,这在很大程度上具有主观性。射线照相只能提供时间上的快照,由于物流和辐射暴露的限制,这些快照是有限的。我们最近提出通过连续监测植入物的负载来评估骨愈合状态,利用植入式传感器系统,即骨折监测器。该设备通过患者的手机以遥测方式传输统计衍生的植入物参数,以帮助医生进行诊断和治疗决策。这项临床前研究旨在系统地研究该设备在动物环境中的安全性和性能。

在 11 只瑞士山地绵羊中进行了不同大小(0.6-30 毫米)的中段胫骨截骨术。骨头用传统的钛或不锈钢锁定板固定,锁定板上装有骨折监测器。在设备的整个使用寿命内连续收集数据。在研究期间,定期进行常规射线照相和临床 CT 扫描。系统地对射线照片进行评分,并对 CT 进行评估以确定缺损处的归一化骨体积。在 9 个月后,对动物进行安乐死。传感器输出与放射学参数相关。从设备位置采集组织样本进行组织学检查。

传感器在能量耗尽之前自主运行了 6.5-8.4 个月。没有观察到设备植入的宏观或微观不良反应。术后 4 周和 8 周的相对植入物负载与射线照相评分和归一化骨体积指标显著相关。

连续的植入物负载监测似乎是一种支持和客观评估骨折愈合的相关方法,并具有在未来实现患者定制康复的强大潜力。