Beat-to-beat blood pressure measurement using a cuffless device does not accurately reflect invasive blood pressure.

作者信息

Moharram Mohammed A, Wilson Luke C, Williams Michael Ja, Coffey Sean

机构信息

Department of Medicine - HeartOtago, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand.

出版信息

Int J Cardiol Hypertens. 2020 Apr 14;5:100030. doi: 10.1016/j.ijchy.2020.100030. eCollection 2020 Jun.

Abstract

BACKGROUND

The availability of an accurate continuous cuffless blood pressure (BP) monitor would provide an alternative to both invasive continuous BP and 24-h intermittent cuff-based BP monitors. We investigated the accuracy of a cuffless beat to beat (BtB) device compared to both invasive BP (iBP) and brachial cuff BP (cBP) measurements.

METHODS

Patients undergoing clinically indicated coronary angiography (CA) and/or percutaneous coronary intervention (PCI) were recruited. After calibration to an initial cBP reading, BP was measured simultaneously using a BtB device (SOMNOtouch NIBP), brachial artery iBP, and cBP at two time points.

RESULTS

The study was terminated early due to a significant bias. Recordings from 14 participants (11 males, mean age 68.4 years) were analysed. Readings from BtB BP were higher than iBP. The bias between BtB BP and iBP was 34.3 mmHg (95%CI: 27.0, 41.5) and 23.6 mmHg (95%CI: 16.8, 30.4) for SBP and DBP respectively. A similar bias was seen between BtB BP and cBP, but cBP and iBP were largely in agreement.

CONCLUSIONS

In patients undergoing CA/PCI, significant differences were detected between BtB BP and both invasively measured and cuff BP. The non-invasive BtB BP measurement device tested is not suitable for clinical or research use.

摘要

背景

准确的连续无袖带血压监测仪的出现,将为有创连续血压监测和24小时间歇性袖带式血压监测仪提供一种替代方案。我们研究了一种无袖带逐搏(BtB)设备与有创血压(iBP)和肱动脉袖带血压(cBP)测量相比的准确性。

方法

招募接受临床指征冠状动脉造影(CA)和/或经皮冠状动脉介入治疗(PCI)的患者。在根据初始cBP读数进行校准后,在两个时间点同时使用BtB设备(SOMNOtouch NIBP)、肱动脉iBP和cBP测量血压。

结果

由于存在显著偏差,该研究提前终止。分析了14名参与者(11名男性,平均年龄68.4岁)的记录。BtB血压读数高于iBP。收缩压和舒张压在BtB血压与iBP之间的偏差分别为34.3 mmHg(95%CI:27.0,41.5)和23.6 mmHg(95%CI:16.8,30.4)。BtB血压与cBP之间也观察到类似偏差,但cBP和iBP基本一致。

结论

在接受CA/PCI的患者中,BtB血压与有创测量血压和袖带血压之间均检测到显著差异。所测试的无创BtB血压测量设备不适合临床或研究使用。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/26f6/7803068/42fd641a1310/gr1.jpg

相似文献

1
Beat-to-beat blood pressure measurement using a cuffless device does not accurately reflect invasive blood pressure.
Int J Cardiol Hypertens. 2020 Apr 14;5:100030. doi: 10.1016/j.ijchy.2020.100030. eCollection 2020 Jun.
2
Comparison of cuff-based and cuffless continuous blood pressure measurements in children and adolescents.
Clin Exp Hypertens. 2020 Aug 17;42(6):512-518. doi: 10.1080/10641963.2020.1714642. Epub 2020 Jan 15.
6
7
Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.
Heart. 2019 Mar;105(5):399-405. doi: 10.1136/heartjnl-2018-313592. Epub 2018 Sep 18.
8
Ability of a 24-h ambulatory cuffless blood pressure monitoring device to track blood pressure changes in clinical practice.
J Hypertens. 2024 Apr 1;42(4):662-671. doi: 10.1097/HJH.0000000000003667. Epub 2024 Jan 23.
9
Practical Application of a New Cuffless Blood Pressure Measurement Method.
Pathophysiology. 2023 Dec 1;30(4):586-598. doi: 10.3390/pathophysiology30040042.

引用本文的文献

1
Accuracy of non-invasive cuffless blood pressure in the intensive care unit: Promises and challenges.
Front Med (Lausanne). 2023 Apr 17;10:1154041. doi: 10.3389/fmed.2023.1154041. eCollection 2023.
2
Measuring blood pressure by a cuffless device using the pulse transit time.
Int J Cardiol Hypertens. 2020 Dec 8;8:100072. doi: 10.1016/j.ijchy.2020.100072. eCollection 2021 Mar.
3
Interpretation of Heart Rate Variability: The Art of Looking Through a Keyhole.
Front Neurosci. 2020 Dec 21;14:609570. doi: 10.3389/fnins.2020.609570. eCollection 2020.

本文引用的文献

1
The use of photoplethysmography for assessing hypertension.
NPJ Digit Med. 2019 Jun 26;2:60. doi: 10.1038/s41746-019-0136-7. eCollection 2019.
2
Brachial and Radial Systolic Blood Pressure Are Not the Same.
Hypertension. 2019 May;73(5):1036-1041. doi: 10.1161/HYPERTENSIONAHA.119.12674.
3
Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.
Heart. 2019 Mar;105(5):399-405. doi: 10.1136/heartjnl-2018-313592. Epub 2018 Sep 18.
4
2018 ESC/ESH Guidelines for the management of arterial hypertension.
Eur Heart J. 2018 Sep 1;39(33):3021-3104. doi: 10.1093/eurheartj/ehy339.
7
A Literature Review on Current and Proposed Technologies of Noninvasive Blood Pressure Measurement.
Telemed J E Health. 2018 Mar;24(3):185-193. doi: 10.1089/tmj.2017.0068. Epub 2017 Aug 7.
9
The Effects of Hemodynamic Changes on Pulse Wave Velocity in Cardiothoracic Surgical Patients.
Biomed Res Int. 2016;2016:9640457. doi: 10.1155/2016/9640457. Epub 2016 Nov 9.
10
New Approaches to Evaluating and Monitoring Blood Pressure.
Curr Hypertens Rep. 2016 Jun;18(6):49. doi: 10.1007/s11906-016-0650-9.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验