新型无袖带设备在临床实践中 24 小时动态血压监测的准确性。

Accuracy of 24-hour ambulatory blood pressure monitoring by a novel cuffless device in clinical practice.

机构信息

Department of Cardiology, University Hospital of Basel, Basel, Switzerland.

Cardiovascular Research Institute Basel, University Hospital of Basel, Basel, Switzerland.

出版信息

Heart. 2019 Mar;105(5):399-405. doi: 10.1136/heartjnl-2018-313592. Epub 2018 Sep 18.

Abstract

OBJECTIVE

Recently, a cuffless blood pressure (BP) measurement device using pulse transit time (PTT) for beat-to-beat calculation of BP values has been validated over a short time period. However, it remains unknown how values obtained with this device compare with standard ambulatory measurements over a 24-hour period. We hypothesised that BP values measured by a cuffless PTT device (TestBP) are comparable with measurements by a standard upper arm cuff-based BP device (RefBP) in clinical practice over 24 hours.

METHODS

Between May and December 2017, 71 individuals were prospectively included. Cuffless using the Somnotouch-NIBP (Somnomedics GmbH, Randersacker, Germany) and cuff-based standard 24-hour BP measurements were performed simultaneously on the left and right arm, respectively. The first RefBP measurement was used as calibration measurement for the TestBP.

RESULTS

Mean (±SD) age was 49.3 (15.1) years, and 51% were male. Mean 24-hour BP for TestBP and RefBP were 140.8 (20.0) versus 134.0 (17.3) mm Hg for systolic (p<0.0001) and 85.8 (14.1)versus79.3 (11.7) mm Hg for diastolic (p<0.0001) measurements, respectively. Mean absolute systolic and diastolic disagreements between TestBP and RefBP were 10.2 (7.2) and 8.2 (5.5) mm Hg, respectively. The number (percentage) of absolute differences between the mean 24-hour BP values of the TestBP and RefBP within 5, 10 and 15 mm Hg were 23 (32.4), 43 (60.6) and 54 (74.6) for systolic and 24 (33.8), 51 (71.8) and 65 (91.6) for diastolic measurements, respectively.

CONCLUSION

In clinical practice over 24 hours, there was a significant difference between the TestBP and RefBP with higher systolic and diastolic BP measured with the cuffless PTT device. Reasons for this difference need to be investigated.

CLINICAL TRIAL REGISTRATION

NCT03054688; Results.

摘要

目的

最近,一种使用脉搏传输时间(PTT)进行逐搏血压值计算的无袖带血压(BP)测量设备已经过短期验证。然而,目前尚不清楚该设备在 24 小时内获得的数值与标准动态血压测量值相比如何。我们假设在 24 小时内,无袖带 PTT 设备(TestBP)测量的 BP 值与标准上臂袖带式 BP 设备(RefBP)的测量值相当。

方法

在 2017 年 5 月至 12 月期间,前瞻性纳入了 71 名个体。分别同时在左臂和右臂上进行无袖带 Somnotouch-NIBP(Somnomedics GmbH,Randersacker,德国)和基于袖带的标准 24 小时 BP 测量。第一次 RefBP 测量被用作 TestBP 的校准测量。

结果

平均(±SD)年龄为 49.3(15.1)岁,51%为男性。TestBP 和 RefBP 的 24 小时平均 BP 分别为收缩压 140.8(20.0)mmHg 对 134.0(17.3)mmHg(p<0.0001)和舒张压 85.8(14.1)mmHg 对 79.3(11.7)mmHg(p<0.0001)。TestBP 和 RefBP 之间的收缩压和舒张压平均绝对差异分别为 10.2(7.2)mmHg 和 8.2(5.5)mmHg。TestBP 和 RefBP 的 24 小时平均 BP 值之间的绝对差值在 5、10 和 15mmHg 以内的数量(百分比)分别为收缩压 23(32.4)、43(60.6)和 54(74.6),舒张压 24(33.8)、51(71.8)和 65(91.6)。

结论

在 24 小时的临床实践中,无袖带 PTT 设备测量的收缩压和舒张压值与 RefBP 之间存在显著差异。需要调查造成这种差异的原因。

临床试验注册号

NCT03054688;结果。