Department of Vascular Surgery, Tianjin First Center Hospital, Tianjin, China.
Department of Pharmacy, The Affiliated Hospital of Qingdao University, Qingdao City, China.
Artif Cells Nanomed Biotechnol. 2021 Dec;49(1):30-37. doi: 10.1080/21691401.2020.1865992.
This research is focussed to quantify IGF1 by electroanalytical analysis on InterDigitated electrode surface and characterized by the microscopic observations. For the detection, antibody and aptamer were used to analyze the level of IGF1. The sandwich pattern (aptamer-IGF1-antibody) was designed on the chemically modified IDE surface and reached the limit of detection to 10 fM with 100 folds enhancement in the sensitivity. Different control experiments (absence of IGF1, binding with IGF2 and with non-complementary aptamer) were failed to show the current changes, discriminated the specific detection. A good detection strategy is to complement the currently following imaging systems for AAA.
本研究旨在通过在叉指电极表面进行电化学生物分析来定量 IGF1,并通过微观观察进行表征。在检测方面,使用抗体和适体来分析 IGF1 的水平。在化学修饰的 IDE 表面上设计了三明治模式(适体-IGF1-抗体),检测限达到 10 fM,灵敏度提高了 100 倍。不同的对照实验(无 IGF1、与 IGF2 结合和与非互补适体结合)均未能显示电流变化,从而区分了特异性检测。一种很好的检测策略是补充当前的 AAA 成像系统。