Suppr超能文献

Sysmex UN 自动化尿液分析仪中的非典型细胞参数:来自现场的反馈。

Atypical cells parameter in Sysmex UN automated urine analyzer: feedback from the field.

机构信息

Central Laboratory, Kepez Public Hospital, Hüsnü Karakaş Mahallesi, No:124, Güneş Cd, 07320, Kepez, Antalya, Turkey.

出版信息

Diagn Pathol. 2021 Jan 22;16(1):9. doi: 10.1186/s13000-021-01068-5.

Abstract

BACKGROUND

"Atypical cells" parameter in automated urinalysis has recently been introduced. An instrument capable of measuring quantitative and qualitative features of nuclear and cytoplasmic properties of a cell has the potential to detect cellular atypia. Instruments using flow cytometry have been detecting atypical cells in blood for a long time; yet instruments using the same methodology very lately developed this parameter in urinalysis.

MATERIALS AND METHODS

Samples with an atypical cells value higher than 1 atypical cell/µL were included in the study. Besides automated urinalysis, every sample was reflexed to modular unit for digital imaging. The remainder of each sample was stained with Sternheimer dye and examined manually under a light microscope.

RESULTS

50 samples with higher than1 atypical cell/µL result were included in the study. Patients were composed of 43 females (86 %) and 7 males (14 %). The mean age was 47.12 ± 19.45 years. The median atypical cells value was 1.8/µL (95 % range 1.5-2.4/µL). Manual microscopic evaluation of the 50 samples showed atypical cells in 1 sample. The patient had papillary lesions on cystoscopy and pathology report informed a high grade urothelial carcinoma. Other 49 samples were negative for atypical cells in manual microscopy. They were crowded samples with leucocytes and squamous epithelial cells.

CONCLUSIONS

The positive case provided evidence for Sysmex UN's capability to detect atypical cells in urine. The negative cases presented clues that probable vulvovaginal contamination and crowded specimens could be deceptive for Sysmex UN in this particular parameter.

摘要

背景

自动化尿液分析中最近引入了“非典型细胞”参数。能够测量细胞核和细胞质特性的定量和定性特征的仪器有可能检测到细胞异型性。使用流式细胞术的仪器已经在血液中检测到非典型细胞很长时间了;然而,使用相同方法学的仪器最近才在尿液分析中开发了这个参数。

材料和方法

将异常细胞值大于 1 个/μL 的样本纳入研究。除了自动尿液分析外,每个样本都被反射到模块化单元进行数字成像。每个样本的其余部分用 Sternheimer 染料染色,并在光显微镜下手动检查。

结果

纳入研究的异常细胞值大于 1 个/μL 的 50 个样本。患者由 43 名女性(86%)和 7 名男性(14%)组成。平均年龄为 47.12±19.45 岁。中位数异常细胞值为 1.8/μL(95%范围 1.5-2.4/μL)。对 50 个样本的手动显微镜评估显示,1 个样本中有非典型细胞。该患者在膀胱镜检查中有乳头状病变,病理报告告知高级别尿路上皮癌。其他 49 个样本在手动显微镜下未见非典型细胞。它们是白细胞和鳞状上皮细胞拥挤的样本。

结论

阳性病例为 Sysmex UN 在尿液中检测非典型细胞的能力提供了证据。阴性病例提示可能的外阴阴道污染和拥挤标本可能会对 Sysmex UN 在这个特定参数中产生误导。