This study aimed to evaluate the property of small dosimeters used for measuring eye lens doses for medical staff during fluoroscopic examination. Dose linearity, energy dependence, and directional dependence of scattered X-rays were evaluated for small radiophotoluminescence glass dosimeters (RPLDs), those with a tin filter (Sn-RPLDs), and small optically stimulated luminescence dosimeters (OSLDs). These dosimeters were pasted on radioprotective glasses, and accumulated air kerma was obtained after irradiating the X-rays to a patient phantom. Strong correlations existed between fluoroscopic time and accumulated air kerma in all types of dosimeters. The energy dependence of Sn-RPLD and OSLD was smaller than that of RPLD. The relative dose value of the OSLD gradually decreased as the angle of the OSLD against the scattered X-rays was larger or lower than the right angle in the horizontal direction. The ranges of relative dose values of RPLD and Sn-RPLD were larger than that of OSLD in the vertical direction. The OSLDs showed lower doses than the RPLDs and Sn-RPLDs, especially on the right side of the radioprotective glasses. These results showed that RPLDs, Sn-RPLDs, and OSLDs had different dosimeter properties, and influence measured eye lens doses for the physician, especially on the opposite side of the patient.
本研究旨在评估用于测量医护人员在荧光透视检查期间眼晶状体剂量的小型剂量计的性能。对小型放射光致发光玻璃剂量计(RPLD)、带锡滤过器的放射光致发光玻璃剂量计(Sn-RPLD)和小型光激励发光剂量计(OSLD)的散射X射线的剂量线性、能量依赖性和方向依赖性进行了评估。将这些剂量计粘贴在辐射防护眼镜上,对患者体模进行X射线照射后获得累积空气比释动能。在所有类型的剂量计中,荧光透视时间与累积空气比释动能之间存在强相关性。Sn-RPLD和OSLD的能量依赖性小于RPLD。在水平方向上,当OSLD相对于散射X射线的角度大于或小于直角时,OSLD的相对剂量值逐渐降低。在垂直方向上,RPLD和Sn-RPLD的相对剂量值范围大于OSLD。OSLD显示的剂量低于RPLD和Sn-RPLD,尤其是在辐射防护眼镜的右侧。这些结果表明,RPLD、Sn-RPLD和OSLD具有不同的剂量计特性,并且会影响医生测量的眼晶状体剂量,尤其是在患者的对侧。