Huizenga H, Levendag P C, De Porre P M, Visser A G
Dr Daniel den Hoed Cancer Center, Rotterdam, The Netherlands.
Radiother Oncol. 1988 Feb;11(2):181-7. doi: 10.1016/0167-8140(88)90254-x.
A prospective study has been performed to determine the accuracy of radiation field alignment for a group of 22 patients with tumors in the head and neck. The accuracy was assessed by an analysis of 138 megavolt portal films in comparison to 55 simulation films. The distance (at the patient midplane) between corresponding points at the field edges on verification film and simulation film appeared to be 5 mm on the average and the standard deviation 5 mm. The analysis was extended by translational and rotational matching of the fields in order to separate each error in a translation error of the field with respect to the patient and an error in field size or shape. Translation errors appear to be somewhat larger than field size or shape errors. From an analysis of a series of megavolt films taken every third radiotherapy session, it was concluded that treatment-to-treatment variations are as large as the errors due to the transition from simulation to treatment situation. Further analysis showed that variation of the patient's position within the cast is clearly one of the error sources.
已开展一项前瞻性研究,以确定22名头颈部肿瘤患者的辐射野对准精度。通过分析138张兆伏级射野片并与55张模拟片进行比较来评估精度。验证片和模拟片上射野边缘对应点之间(在患者中平面处)的距离平均似乎为5毫米,标准差为5毫米。通过对射野进行平移和旋转匹配来扩展分析,以便区分射野相对于患者的平移误差以及射野大小或形状的误差。平移误差似乎比射野大小或形状误差略大。通过对每第三次放疗疗程拍摄的一系列兆伏级片子进行分析得出结论,治疗间的差异与从模拟到治疗情况转变所导致的误差一样大。进一步分析表明,患者在模型内位置的变化显然是误差来源之一。