Abstract

BACKGROUND

Accurate serum creatinine (Cr) concentration measurement is essential for evaluating kidney function. In 2011, the Korean Association of External Quality Assessment Service (KEQAS) launched an accuracy-based Cr proficiency testing (ABCr PT) survey. We analyzed long-term data of the KEQAS ABCr PT survey collected between 2011 and 2019 to assess recent trends in Cr assays in Korea.

METHODS

The ABCr PT survey including three commutable fresh-frozen serum samples was performed twice a year. The target Cr concentration was assigned using isotope-dilution mass spectrometry. We analyzed data obtained from the participating laboratories, calculated the yearly bias, and evaluated bias trends for the major reagents and instruments. Outliers were excluded from all analysis.

RESULTS

The mean percentage bias based on the total data of all participating laboratories was 10.8% in the 2011-A survey and 0.2% in 2019-B survey. Bias for the major reagents and instruments differed depending on the manufacturer. Enzymatic assays generally showed desirable bias ranging from -3.9% to 3.2% at all Cr concentrations and lower interlaboratory variability than non-enzymatic assays (enzymatic vs. non-enzymatic, 3.3%-7.2% vs. 6.3%-9.1%).

CONCLUSIONS

Although the mean percentage bias of Cr assays tends to decrease over time, it is necessary to continuously strive to improve Cr assay accuracy, especially at low concentrations.

摘要

背景

准确的血清肌酐(Cr)浓度测量对于评估肾功能至关重要。2011 年,韩国外部质量评估服务协会(KEQAS)启动了一项基于准确度的 Cr 能力验证(ABCr PT)调查。我们分析了 2011 年至 2019 年期间收集的 KEQAS ABCr PT 调查的长期数据,以评估韩国 Cr 分析的最新趋势。

方法

ABCr PT 调查包括三个可互换的新鲜冷冻血清样本,每年进行两次。使用同位素稀释质谱法分配目标 Cr 浓度。我们分析了来自参与实验室的数据,计算了每年的偏差,并评估了主要试剂和仪器的偏差趋势。所有分析均排除了离群值。

结果

根据所有参与实验室的总数据,2011-A 调查的平均百分比偏差为 10.8%,2019-B 调查的平均百分比偏差为 0.2%。主要试剂和仪器的偏差因制造商而异。在所有 Cr 浓度下,酶法测定通常显示出理想的偏差范围为-3.9%至 3.2%,且与非酶法测定相比,实验室间变异性较低(酶法与非酶法,3.3%-7.2%与 6.3%-9.1%)。

结论

尽管 Cr 分析的平均百分比偏差随着时间的推移趋于降低,但仍有必要不断努力提高 Cr 分析的准确性,尤其是在低浓度时。