Measurement of arterial pressure using catheter-transducer systems. Improvement using the Accudynamic.

作者信息

Allan M W, Gray W M, Asbury A J

机构信息

Department of Anaesthetics, Western Infirmary, Glasgow.

出版信息

Br J Anaesth. 1988 Mar;60(4):413-8. doi: 10.1093/bja/60.4.413.

Abstract

The dynamic performance of arterial manometers depends upon the values of the natural frequency fo and the damping factor beta. Satisfactory reproduction of all arterial waveforms requires that fo should be greater than about 13 Hz; for an fo of 13 Hz, beta should have a value of 0.5, but, as fo increases, an increasing range of beta values is permissible. We measured fo and beta for two pressure transducers (Gould P23 Statham and Gould Disposable) in combination with Portex or Gould catheter tubing of lengths varying from 100 cm to 300 cm. All combinations were considerably underdamped, with the best having an fo of 30 Hz and a beta of 0.1. We also investigated the performance of a commercial device, the Accudynamic, designed to allow beta to be increased without affecting fo. We found that the Accudynamic worked well, allowing acceptable performance to be obtained from the Gould Disposable transducer with catheters up to 200 cm in length.

摘要

动脉压力计的动态性能取决于固有频率fo和阻尼因子β的值。要令人满意地再现所有动脉波形,要求fo应大于约13Hz;对于13Hz的fo,β应具有0.5的值,但是,随着fo的增加,β值的允许范围也会增加。我们测量了两个压力传感器(古尔德P23斯塔瑟姆和古尔德一次性传感器)与长度从100cm到300cm不等的波特克斯或古尔德导管连接时的fo和β。所有组合的阻尼都明显不足,最好的组合固有频率为30Hz,阻尼因子为0.1。我们还研究了一种商业设备Accudynamic的性能,该设备旨在在不影响fo的情况下增加β。我们发现Accudynamic效果良好,使用长达200cm的导管时,能从古尔德一次性传感器获得可接受的性能。