Hunziker P
Can J Anaesth. 1987 Jul;34(4):409-14. doi: 10.1007/BF03010146.
Accuracy and dynamic response of seven different disposable saline-filled transducer-tubing systems were evaluated in two configurations; a pole-mount version (60-inch tubing) and patient-mount version (12-inch tubing). Natural frequencies (Fn) and damping coefficients were measured as were pressure waveform recordings. A 0.03 ml air bubble was inserted into the different systems in close proximity to the transducer and measurements were repeated. There were dramatic changes in dynamic response with lengthening of tubing to a pole-mount configuration and air bubble additions. Only one transducer-tubing system had adequate accuracy and dynamic response in both the pole- and patient-mount version. Pressure waveforms in the pole-mount versions and with the air bubble produced systolic overshoots up to 35 mmHg. The performance characteristics dictate that the choice for a disposable transducer-tubing system must be made not only by design, but dynamic response which is directly related to clinical set-up.
评估了七种不同的一次性充生理盐水换能器-管道系统在两种配置下的准确性和动态响应;一种是杆装式版本(60英寸管道)和患者装式版本(12英寸管道)。测量了固有频率(Fn)和阻尼系数以及压力波形记录。在靠近换能器的不同系统中插入了一个0.03毫升的气泡,并重复进行测量。随着管道延长至杆装式配置以及添加气泡,动态响应发生了显著变化。只有一个换能器-管道系统在杆装式和患者装式版本中都具有足够的准确性和动态响应。杆装式版本以及有气泡时的压力波形产生了高达35 mmHg的收缩期过冲。性能特征表明,一次性换能器-管道系统的选择不仅要基于设计,还要考虑与临床设置直接相关的动态响应。