Suppr超能文献

血压监测系统快速冲洗测试与方波测试的等效性。

Equivalence of fast flush and square wave testing of blood pressure monitoring systems.

作者信息

Kleinman B, Powell S, Gardner R M

机构信息

Department of Anesthesia, Loyola University Medical Center, Maywood, Illinois 60153, USA.

出版信息

J Clin Monit. 1996 Mar;12(2):149-54. doi: 10.1007/BF02078135.

Abstract

BACKGROUND

The accurate recording of intraarterial pressure depends upon an appropriate dynamic response of the monitoring system. Generation of a square wave (SW) at the catheter tip is the engineering and in vitro laboratory gold standard. Fast flush (FF) testing is the clinical test of choice. Results from these two test methods have been assumed equal but have not been empirically confirmed.

METHODS

We studied three different 5.1 cm catheter sizes (16 G, 18 G, 20 G Becton Dickinson, Sandy, UT) attached to three different lengths of arterial pressure tubing (36 in, 91.4 cm; 72 in, 182.9 cm; 108 in, 274.3 cm). An arterial recording system was assembled in the standard fashion by attaching a catheter to arterial pressure tubing, which was attached to a transducer (TXX-R, Ohmeda, formerly Viggo-Spectramed, Oxnard, CA) whose signal was recorded by a strip chart recorder (Gould 2400, Rolling Meadows, IL). The system was attached to a pressurized saline flush. The catheter tip was inserted into one port of a pressure generator. With the other port of the pressure generator open to atmosphere, FF tests were performed by activating the flush device of the transducer. Subsequent step response signals from the FF tests were then recorded from which natural frequency (fn) and damping coefficient (zeta) were calculated. Next, square waves were generated by closing the port that was open to atmosphere and attaching a signal generator to a pressure generator. Square waves so generated were recorded as described above and natural frequency and damping coefficients calculated. These procedures were repeated after 0.05 cc of air was introduced in the transducer and repeated again in a system containing a damping device (R.O.S.E., Resonant OverShoot Eliminator, Viggo-Spectramed, Oxnard, CA).

RESULTS

There was no significant difference between fn and zeta as calculated from the step response generated from the FF test versus fn and zeta as calculated from the square wave (SW) test in systems without air. However, in systems containing air, fn by FF testing was always less than fn by SW testing for all catheter sizes and extension tubing lengths (p < 0.05). Damping was also always greater by FF testing than by SW testing in systems with air for all catheter sizes and extension tubing lengths (p < 0.05). The R.O.S.E device created marked qualitative differences, although exact fn and zeta could not be quantified.

CONCLUSIONS

For the characterization of dynamic response of invasive blood pressure monitoring systems, the FF test and SW test yield identical results. However, under certain conditions-air, R.O.S.E device-dynamic response as measured by FF testing was not equivalent to dynamic response as measured by the gold standard-the SW test. Specifically, small amounts of air in fluid-filled invasive blood pressure monitoring systems cause a slightly worse dynamic response as measured by FF testing versus the laboratory gold standard-the SW test.

摘要

背景

动脉内压力的准确记录取决于监测系统适当的动态响应。在导管尖端产生方波(SW)是工程和体外实验室的金标准。快速冲洗(FF)测试是临床首选的测试方法。这两种测试方法的结果一直被认为是相等的,但尚未得到经验证实。

方法

我们研究了连接到三种不同长度动脉压管(36英寸,91.4厘米;72英寸,182.9厘米;108英寸,274.3厘米)的三种不同尺寸(16G、18G、20G,美国犹他州桑迪市贝克顿·迪金森公司生产)的5.1厘米导管。通过将导管连接到动脉压管,再将动脉压管连接到传感器(TXX - R,奥米达公司,原维戈 - 斯佩克特拉姆公司,加利福尼亚州奥克斯纳德市),以标准方式组装动脉记录系统,传感器的信号由带状图表记录仪(古尔德2400,伊利诺伊州罗灵梅多斯市)记录。该系统连接到加压盐水冲洗装置。将导管尖端插入压力发生器的一个端口。在压力发生器的另一个端口通向大气的情况下,通过启动传感器的冲洗装置进行FF测试。然后记录FF测试的后续阶跃响应信号,并据此计算固有频率(fn)和阻尼系数(ζ)。接下来,关闭通向大气的端口并将信号发生器连接到压力发生器,产生方波。如此产生的方波按上述方法记录,并计算固有频率和阻尼系数。在传感器中引入0.05毫升空气后重复这些步骤,并在包含阻尼装置(R.O.S.E.,共振过冲消除器,维戈 - 斯佩克特拉姆公司,加利福尼亚州奥克斯纳德市)的系统中再次重复。

结果

在无空气的系统中,根据FF测试产生的阶跃响应计算得到的fn和ζ与根据方波(SW)测试计算得到的fn和ζ之间无显著差异。然而,在含有空气的系统中,对于所有导管尺寸和延长管长度,通过FF测试得到的fn总是小于通过SW测试得到的fn(p < 0.05)。在含有空气的系统中,对于所有导管尺寸和延长管长度,通过FF测试得到的阻尼也总是大于通过SW测试得到的阻尼(p < 0.05)。尽管无法精确量化确切的fn和ζ,但R.O.S.E.装置产生了明显的定性差异。

结论

对于有创血压监测系统动态响应的表征,FF测试和SW测试产生相同的结果。然而,在某些条件下——空气、R.O.S.E.装置——通过FF测试测量的动态响应与通过金标准——SW测试测量的动态响应不相等。具体而言,在充满液体的有创血压监测系统中存在少量空气时,与实验室金标准——SW测试相比,通过FF测试测量的动态响应略差。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验