Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, People's Republic of China.
Beijing Center for Clinical Laboratories, Beijing, People's Republic of China.
Scand J Clin Lab Invest. 2021 Apr;81(2):85-91. doi: 10.1080/00365513.2020.1855469. Epub 2021 Feb 11.
Electrolytes for sodium, potassium, magnesium, and calcium are important serum ions that are frequently assayed in clinical laboratories. In this study, we assessed the trueness of routine analytical systems for four cations using an inexpensive candidate reference method aimed to promote the standardization of serum electrolyte detection. An ion chromatography (IC) method with Cesium as an internal standard was developed and evaluated. The residual clinical serum samples at Chaoyang Hospital were collected and prepared into three human serum pools of electrolytes, which were used for the trueness evaluation of five routine analytical systems. Furthermore, the agreement between routine methods and the IC method was verified using 40 individual human samples. The recovery rates of sodium, potassium, magnesium and calcium were 99.69%, 100.34%, 100.43% and 99.89%, respectively. The intra-batch standard deviation and intra-laboratory precision of NIST SRM 956c were all less than 1% for the four ions. The certified values were within the validation range, and the deviation between the results and the certified values were less than 0.5%. The three serum pools were homogeneous and stable. All routine systems aligned with the IC method for four cations and achieved the analytical quality specifications for potassium and magnesium at 3 different concentrations. The developed IC method is simple, practical, accurate, and precise, which can be used as a candidate reference method for serum electrolytes measurement. Five routine analytical systems for electrolytes measurement had the acceptable bias for potassium and magnesium and their results showed good concordance.
电解质(钠、钾、镁和钙)是血清中的重要离子,在临床实验室中经常进行检测。在这项研究中,我们使用一种廉价的候选参考方法评估了四个阳离子常规分析系统的准确度,旨在促进血清电解质检测的标准化。开发并评估了一种以铯为内标物的离子色谱(IC)方法。收集了朝阳医院的剩余临床血清样本并制备成三个电解质人血清池,用于评估五个常规分析系统的准确度。此外,使用 40 个人类样本验证了常规方法与 IC 方法之间的一致性。钠、钾、镁和钙的回收率分别为 99.69%、100.34%、100.43%和 99.89%。对于这四个离子,NIST SRM 956c 的批内标准差和实验室内精密度均小于 1%。认证值在验证范围内,结果与认证值的偏差小于 0.5%。三个血清池均均匀且稳定。所有常规系统均与 IC 法对于四个阳离子一致,并在 3 种不同浓度下达到了钾和镁的分析质量规格。所开发的 IC 方法简单、实用、准确、精密,可作为血清电解质测量的候选参考方法。用于电解质测量的五个常规分析系统对于钾和镁具有可接受的偏差,并且它们的结果具有良好的一致性。