Abstract

OBJECTIVES

Germanium-68/gallium-68 (Ge/Ga) generator eluate contains a number of metal cations that can compete with GaCl, reducing specific radioactivity. The first step in peptide labeling with GaCl is to remove Ge and several other metals with a long half-life. In this purification step, the elution residue that is passed through the cartridge is collected in glass waste bottles. Waste management is included in good production practices, and in particular, the activity of long half-life Ge (270.95 days) and other toxic metal levels need to be examined. Our objective in this study is to determine the Ge activity in liquid waste produced by the generation of Ga and heavy metal concentrations from the generator column materials and to assess whether it can be disposed of as normal waste.

METHODS

Liquid wastes produced by passing the Ge/Ga generator eluate of 2 different identities via PSH+ cartridge have been analyzed with the inductively coupled plasma mass spectrometry device in the advanced technology application and research center of our university.

RESULTS

The average of the Ge radioactive pollution was estimated to be 0.142 ppm (μg.mL) in the liquid waste analysis after passing through the PSH+ cartridge in the pre-elution in the GalluGEN brand generator. While there was no tin (Sn) impurity, it was determined that the average zinc (Zn) was 1.95 ppm (μg.mL) and the average aluminum (Al) impurity was 10.95 ppm (μg.mL). While no Ge radioactive pollution was determined in the iThemba LABS brand generator, the average Sn was 0.098 ppm (μg.mL), average Zn 48.6 ppm (μg.mL), and average Al impurity 4.135 ppm (μg.mL).

CONCLUSION

All Ge/Ga generators produced have their own certificates. Metallic contamination in the postmarking waste of Ge/Ga generators can be different. It would be a safe method to keep these wastes in place until they are dumped into the sewage systems, given their half-lives in terms of long half-life radioactive metallic contamination.

摘要

目的

锗 - 68/镓 - 68(Ge/Ga)发生器洗脱液含有多种金属阳离子,它们可与氯化镓竞争,降低比活度。用氯化镓进行肽标记的第一步是去除锗和其他几种半衰期长的金属。在该纯化步骤中,通过柱的洗脱残余物收集在玻璃废液瓶中。废物管理包含在良好生产规范中,特别是需要检测半衰期长的锗(270.95天)的活度和其他有毒金属水平。本研究的目的是测定由发生器柱材料产生镓时产生的废液中的锗活度以及重金属浓度,并评估其是否可作为普通废物处置。

方法

通过我校先进技术应用与研究中心的电感耦合等离子体质谱仪,对两种不同标识的Ge/Ga发生器洗脱液通过PSH +柱后产生的废液进行了分析。

结果

在GalluGEN品牌发生器预洗脱过程中,通过PSH +柱后的废液分析中,锗放射性污染的平均值估计为0.142 ppm(μg/mL)。虽然没有锡(Sn)杂质,但确定锌(Zn)的平均值为1.95 ppm(μg/mL),铝(Al)杂质的平均值为10.95 ppm(μg/mL)。在iThemba LABS品牌发生器中未测定到锗放射性污染,锡的平均值为0.098 ppm(μg/mL),锌的平均值为48.6 ppm(μg/mL),铝杂质的平均值为4.135 ppm(μg/mL)。

结论

所有生产的Ge/Ga发生器都有其自身的证书。Ge/Ga发生器标记后废物中的金属污染可能不同。鉴于其在长半衰期放射性金属污染方面的半衰期,在将这些废物倒入污水系统之前将其妥善保存是一种安全的方法。