Medical Physics and Research Department, Hong Kong Sanatorium and Hospital, Hong Kong SAR, People's Republic of China.
Phys Med Biol. 2021 Mar 12;66(6):065021. doi: 10.1088/1361-6560/abe837.
Bolus is commonly used in MV photon radiotherapy to increase superficial dose and improve dose uniformity for treating shallow lesions. However, irregular patient body contours can cause unwanted air gaps between a bolus and patient skin. The resulting dosimetric errors could be exacerbated in MR-Linac treatments, as secondary electrons generated by photons are affected by the magnetic field. This study aimed to quantify the dosimetric effect of unwanted gaps between bolus and skin surface in an MR-Linac. A parallel-plate ionization chamber and EBT3 films were utilized to evaluate the surface dose under bolus with various gantry angles, field sizes, and different air gaps. The results of surface dose measurements were then compared to Monaco 5.40 Treatment Planning System (TPS) calculations. The suitability of using a parallel-plate chamber in MR-Linac measurement was validated by benchmarking the percentage depth dose and output factors with the microDiamond detector and air-filled ionization chamber measurements in water. A non-symmetric response of the parallel-plate chamber to oblique beams in the magnetic field was characterized. Unwanted air gaps significantly reduced the skin dose. For a frontal beam, skin dose was halved when there was a 5 mm gap, a much larger difference than in a conventional linac. Skin dose manifested a non-symmetric pattern in terms of gantry angle and gap size. The TPS overestimated skin dose in general, but shared the same trend with measurement when there was no air gap, or the gap size was larger than 5 mm. However, the calculated and measured results had a large discrepancy when the bolus-skin gap was below 5 mm. When treating superficial lesions, unwanted air gaps under the bolus will compromise the dosimetric goals. Our results highlight the importance of avoiding air gaps between bolus and skin when treating superficial lesions using an MR-Linac system.
团注通常用于 MV 光子放射治疗中,以增加浅层剂量并提高治疗浅层病变的剂量均匀性。然而,不规则的患者体廓可能导致在团注和患者皮肤之间产生不希望的气隙。在 MR-Linac 治疗中,光子产生的次级电子会受到磁场的影响,从而导致更大的剂量学误差。本研究旨在定量评估 MR-Linac 中团注和皮肤表面之间不希望的气隙的剂量学影响。使用平行板电离室和 EBT3 胶片评估了在不同机架角度、射野大小和不同气隙下的表面剂量。然后将表面剂量测量结果与 Monaco 5.40 治疗计划系统(TPS)计算结果进行比较。通过将百分深度剂量和输出因子与微钻石探测器和充满空气的电离室在水中的测量值进行基准测试,验证了平行板电离室在 MR-Linac 测量中的适用性。对平行板电离室在磁场中对斜束的非对称响应进行了表征。不希望的气隙会显著降低皮肤剂量。对于前向射束,当气隙为 5mm 时,皮肤剂量减半,这比在常规直线加速器中要大得多。皮肤剂量在机架角度和气隙大小方面表现出非对称模式。TPS 通常高估了皮肤剂量,但在没有气隙或气隙尺寸大于 5mm 时,与测量值具有相同的趋势。然而,当团注-皮肤间隙小于 5mm 时,计算和测量结果之间存在较大差异。在治疗浅层病变时,团注下不希望的气隙会影响剂量学目标。我们的研究结果强调了在使用 MR-Linac 系统治疗浅层病变时避免团注和皮肤之间气隙的重要性。