Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI, USA.
J Appl Clin Med Phys. 2019 Jul;20(7):28-38. doi: 10.1002/acm2.12651. Epub 2019 Jun 29.
The magnetic field can cause a nonnegligible dosimetric effect in an MR-Linac system. This effect should be accurately accounted for by the beam models in treatment planning systems (TPS). The purpose of the study was to verify the beam model and the entire treatment planning and delivery process for a 1.5 T MR-Linac based on comprehensive dosimetric measurements and end-to-end tests.
Dosimetry measurements and end-to-end tests were performed on a preclinical MR-Linac (Elekta AB) using a multitude of detectors and were compared to the corresponding beam model calculations from the TPS for the MR-Linac. Measurement devices included ion chambers (IC), diamond detector, radiochromic film, and MR-compatible ion chamber array and diode array. The dose in inhomogeneous phantom was also verified. The end-to-end tests include the generation, delivery, and comparison of 3D and IMRT plan with measurement.
For the depth dose measurements with Farmer IC, micro IC and diamond detector, the absolute difference between most measurement points and beam model calculation beyond the buildup region were <1%, at most 2% for a few measurement points. For the beam profile measurements, the absolute differences were no more than 1% outside the penumbra region and no more than 2.5% inside the penumbra region. Results of end-to-end tests demonstrated that three 3D static plans with single 5 × 10 cm fields (at gantry angle 0°, 90° and 270°) and two IMRT plans successfully passed gamma analysis with clinical criteria. The dose difference in the inhomogeneous phantom between the calculation and measurement was within 1.0%.
Both relative and absolute dosimetry measurements agreed well with the TPS calculation, indicating that the beam model for MR-Linac properly accounts for the magnetic field effect. The end-to-end tests verified the entire treatment planning process.
磁场会在 MR-Linac 系统中引起不可忽略的剂量学效应。这种效应应通过治疗计划系统 (TPS) 中的射束模型准确考虑。本研究的目的是基于全面的剂量测量和端到端测试,验证 1.5T MR-Linac 的射束模型和整个治疗计划与交付过程。
使用多种探测器在临床前 MR-Linac(Elekta AB)上进行剂量学测量和端到端测试,并将其与 MR-Linac 的 TPS 相应的射束模型计算进行比较。测量设备包括电离室 (IC)、钻石探测器、光致变色胶片、MR 兼容电离室阵列和二极管阵列。还验证了不均匀体模中的剂量。端到端测试包括 3D 和调强放疗计划的生成、交付和比较,以及测量。
对于 Farmer IC、微 IC 和钻石探测器的深度剂量测量,大多数测量点与累积区外射束模型计算的绝对差值<1%,少数测量点的差值最大为 2%。对于射束轮廓测量,在半影区外的绝对差值不超过 1%,在半影区内的绝对差值不超过 2.5%。端到端测试的结果表明,三个具有单 5×10cm 野(在机架角度 0°、90°和 270°)的 3D 静态计划和两个 IMRT 计划均成功通过了临床标准的伽马分析。计算与测量之间不均匀体模中的剂量差异在 1.0%以内。
相对和绝对剂量学测量与 TPS 计算吻合良好,表明 MR-Linac 的射束模型正确考虑了磁场效应。端到端测试验证了整个治疗计划过程。