Associates in Medical Physics (AMP), Greenbelt, MD, USA.
Department of Radiation Medicine, Oregon Health and Science University, Portland, OR, USA.
J Appl Clin Med Phys. 2021 Mar;22(3):72-85. doi: 10.1002/acm2.13180. Epub 2021 Feb 22.
In a mixed-vendor radiation oncology environment, it is advantageous if the department's treatment planning system (TPS) supports the linear accelerators of different vendors. In this publication beam data collection and modeling for the Versa HD linear accelerator in Monaco, Pinnacle, and Eclipse are discussed. In each TPS static field, Intensity-Modulated Radiation Therapy (IMRT) step and shoot, and Volumetric-Modulated Arc Therapy (VMAT) plans for flattened and flattening-filter free photon beams of all available energies were evaluated for field sizes >3 × 3. To compare passing rates, identical beam model validation plans were calculated in each TPS. Eclipse, Monaco, and Pinnacle beam models passed validation measurements in homogeneous materials for a variety of treatment fields, including static, IMRT, and VMAT. In the case of Eclipse, the "dosimetric leaf gap" parameter was found to be critical for passing rates of VMAT plans. The source size parameter plays an important role as well for small fields. In the case of Pinnacle the multileaf collimator offset table needed to be optimized for better VMAT QA results. Each of the investigated treatment planning systems met the criteria to be used clinically in conjunction with Elekta Versa HD linear accelerators. It can be of great advantage to have the option to operate a TPS and linear accelerator from different vendors, as decisions surrounding linear accelerator or TPS purchases are very complicated and not just limited to technical considerations.
在混合供应商放射肿瘤学环境中,如果部门的治疗计划系统 (TPS) 支持不同供应商的线性加速器,这将是有利的。在本出版物中,讨论了 Monaco、Pinnacle 和 Eclipse 中的 Versa HD 线性加速器的束数据采集和建模。在每个 TPS 中,对所有可用能量的平化和非平化滤过光子束的静态场、调强放射治疗 (IMRT) 步进和拍摄以及容积调强弧形治疗 (VMAT) 进行了评估,场大小 >3×3。为了比较通过率,在每个 TPS 中计算了相同的束模型验证计划。Eclipse、Monaco 和 Pinnacle 的束模型在各种治疗场,包括静态、IMRT 和 VMAT 的均匀材料中通过了验证测量。在 Eclipse 的情况下,“剂量学叶片间隙”参数对于 VMAT 计划的通过率至关重要。对于小场,源尺寸参数也起着重要作用。在 Pinnacle 的情况下,需要优化多叶准直器偏移表,以获得更好的 VMAT QA 结果。所研究的每个治疗计划系统都符合与 Elekta Versa HD 线性加速器一起临床使用的标准。能够选择操作来自不同供应商的 TPS 和线性加速器是非常有利的,因为围绕线性加速器或 TPS 购买的决策非常复杂,不仅仅局限于技术考虑。