Department of Radiation Oncology, Mayo Clinic Scottsdale, 5777 East Mayo Boulevard, MCSB Concourse, Phoenix, Arizona 89054, USA.
Med Phys. 2009 Nov;36(11):5359-73. doi: 10.1118/1.3238104.
AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.
AAPM 工作组 119 已经制定了定量置信限作为 IMRT 调试的基线预期值。开发了一组测试用例来评估 IMRT 治疗的计划和交付的整体准确性。每个测试都使用矩形体模内绘制的目标和避让结构的轮廓。这些测试由九家使用各种 IMRT 计划和交付系统的机构进行规划、交付、测量和分析。每个机构都通过了放射物理中心的 IMRT 认证测试。使用高剂量区和低剂量区的电离室剂量学、在有所有射野交付的体模的冠状面上的胶片剂量学以及垂直于中心轴测量的每个射野的平面剂量学,确定计划剂量和测量剂量之间的一致性。使用 3%/3mm 的伽马标准评估平面剂量分布。使用置信限 = /均值/ + 1.96sigma 的概念,使用平均值和标准差为测试结果制定置信限。其他机构可以使用测试协议和结果作为与该组进行比较的基础。本地得出的置信限如果大大超过这些基线值,则可能表明需要改进 IMRT 调试。