Electro-Medical Device Research Center, Korea Electrotechnology Research Institute, Ansan 15588, South Korea.
RSS Center, Korea Electrotechnology Research Institute, Ansan 15588, South Korea.
Rev Sci Instrum. 2021 Feb 1;92(2):024103. doi: 10.1063/5.0030271.
A compact X-band linear accelerator (LINAC) system equipped with a small and lightweight magnetron was constructed to develop a high-precision image-guided radiotherapy system. The developed LINAC system was installed in an O-ring gantry where cone-beam computed tomography (CBCT) was embedded. When the O-arm gantry is rotated, an x-ray beam is stably generated, which resulted from the stable transmission of radio frequency power into the X-band LINAC system. Quality assurance (QA) tests, including mechanical and dosimetry checks, were carried out to ensure safety and operation performance according to the American Association of Physicists in Medicine's TG-51, 142, an international standard protocol established by accredited institutions. In addition, delivery QA of the radiotherapy planning system was conducted to verify intensity-modulated radiotherapy techniques. Therefore, it was demonstrated that the developed X-band LINAC system mounted on the O-arm gantry proved to be valid and reliable for potential use in CBCT image-guided radiation therapy.
构建了配备小型轻量微波电子管的紧凑型 X 波段直线加速器 (LINAC) 系统,以开发高精度图像引导放射治疗系统。所开发的 LINAC 系统安装在嵌入了锥形束计算机断层扫描 (CBCT) 的 O 型环龙门架中。当 O 臂龙门架旋转时,由于射频功率稳定地传输到 X 波段 LINAC 系统,因此会稳定地产生 X 射线束。根据美国医学物理学家协会的 TG-51、142 标准,进行了包括机械和剂量学检查在内的质量保证 (QA) 测试,以确保安全和运行性能,该标准是由认证机构制定的国际标准协议。此外,还进行了放射治疗计划系统的交付 QA 测试,以验证强度调制放射治疗技术。因此,证明了安装在 O 臂龙门架上的开发的 X 波段 LINAC 系统可有效可靠地用于 CBCT 图像引导放射治疗。