To evaluate the feasibility of Quantitative Ultrashort-Time-to-Echo Contrast-Enhanced (QUTE-CE) MRA using ferumoxytol as a contrast agent for abdominal angiography in the kidney.
Four subjects underwent ferumoxytol-enhanced MRA with the 3D UTE Spiral VIBE WIP sequence at 3 T. Image quality metrics were quantified, specifically the blood Signal-to-Noise Ratio (SNR), blood-tissue Contrast-to-Noise Ratio (CNR) and Intraluminal Signal Heterogeneity (ISH) from both the aorta and inferior vena cava (IVC). Morphometric analysis of the vessels was performed using manual approach and semi-automatic approach using Vascular Modeling ToolKit (VMTK). Image quality and branching order were compared between QUTE-CE MRA and the Gadolinium (Gd) CEMRA reference image.
QUTE-CE MRA provides a bright blood snapshot that delineates arteries and veins equally in the same scan. The maximum SNR and CNR values were 3,282 ± 1,218 and 1,295 ± 580, respectively - significantly higher than available literature values using other CEMRA techniques. QUTE-CE MRA had lower ISH and depicted higher vessel branching order (7th vs 3rd) within the kidney compared to a standard dynamic clinical Gd CEMRA scan. Morphometric analysis yielded quantitative results for the total kidney volume, total cyst volume and for diameters of the branching arterial network down to the 7 branch. Vessel curvature was significantly increased (p < 0.001) in the presence of a renal cyst compared to equivalent vessels in normal kidney regions.
QUTE-CE MRA is feasible for kidney angiography, providing greater detail of kidney vasculature, enabling quantitative morphometric analysis of the abdominal and intra-renal vessels and yielding metrics relevant to vascular diseases while using a contrast agent ferumoxytol that is safe for CKD patients.
评估使用 Ferumoxytol 作为对比剂进行腹部血管造影的定量超短回波时间对比增强(QUTE-CE)MRA 在肾脏中的可行性。
4 名受检者在 3T 下使用 3D UTE Spiral VIBE WIP 序列进行 Ferumoxytol 增强 MRA。量化了图像质量指标,具体包括来自主动脉和下腔静脉(IVC)的血液信噪比(SNR)、血液-组织对比度噪声比(CNR)和管腔内信号异质性(ISH)。使用手动方法和使用血管建模工具包(Vascular Modeling ToolKit,VMTK)的半自动方法对血管进行形态测量分析。比较了 QUTE-CE MRA 和钆(Gd)CE-MRA 参考图像的图像质量和分支顺序。
QUTE-CE MRA 提供了一个明亮的血液快照,在同一扫描中同等程度地描绘了动脉和静脉。最大 SNR 和 CNR 值分别为 3282±1218 和 1295±580,明显高于使用其他 CEMRA 技术的现有文献值。与标准动态临床 Gd CEMRA 扫描相比,QUTE-CE MRA 的 ISH 更低,并且在肾脏内描绘的血管分支顺序更高(第 7 级与第 3 级)。形态测量分析产生了总肾体积、总囊肿体积以及分支动脉网络直径的定量结果,直至第 7 级分支。与正常肾区的等效血管相比,存在肾囊肿时血管曲率明显增加(p<0.001)。
QUTE-CE MRA 可用于肾脏血管造影,提供更详细的肾脏血管图像,实现腹部和肾内血管的定量形态测量分析,并产生与血管疾病相关的指标,同时使用 Ferumoxytol 作为对比剂,对 CKD 患者安全。