Retina. 2021 Oct 1;41(10):2059-2065. doi: 10.1097/IAE.0000000000003163.
To assess the full-thickness macular hole (FTMH) size using the choroidal hypertransmission signal on spectral-domain optical coherence tomography and to compare this method to the standard aperture measurement of the minimum aperture size at the level of the neurosensory retina.
Cross-sectional study of retrospective data.
Eyes with FTMH imaged on spectral-domain optical coherence tomography were included. Two independent masked graders used the device's built-in caliper tool to measure the FTMH minimum aperture size at the level of the neurosensory retina and the size of the corresponding hypertransmission signal below the level of the retinal pigment epithelium/Bruch membrane complex. To assess the reproducibility of the hypertransmission measurement in tilted scans, two measurements were obtained and compared; the first was traced parallel to the retinal pigment epithelium (parallel hypertransmission), and the second was horizontal to the image frame (horizontal hypertransmission), both using Image J software.
A total of 31 eyes were enrolled. The mean FTMH minimum aperture size was smaller compared with both the choroidal parallel hypertransmission and horizontal hypertransmission measurements (mean ± SD: 335.7 ± 139.5 µm, 376.7 ± 150.6 µm, 375.1 ± 150.0 µm, respectively. P < 0.001 for both comparisons).
The proposed hypertransmission measurement is a feasible and reproducible alternative to assess FTMH size and could provide the basis for an automated FTMH measurement on cross-sectional spectral-domain optical coherence tomography scans, as presented in this study, or on the spectral-domain optical coherence tomography volumetric data set by using an en face projection.
利用谱域光相干断层扫描中的脉络膜高透过信号评估全层黄斑裂孔(FTMH)的大小,并将该方法与神经感觉视网膜水平最小孔径的标准孔径测量进行比较。
回顾性数据的横断面研究。
纳入在谱域光相干断层扫描上成像的 FTMH 眼。两名独立的掩蔽评分员使用设备内置的卡尺工具测量神经感觉视网膜水平的 FTMH 最小孔径大小,以及视网膜色素上皮/布鲁赫膜复合体下方相应的高透过信号大小。为了评估倾斜扫描中高透过测量的可重复性,获得并比较了两次测量;第一次是沿视网膜色素上皮平行追踪(平行高透过),第二次是与图像框架水平追踪(水平高透过),均使用 Image J 软件。
共纳入 31 只眼。FTMH 最小孔径大小与脉络膜平行高透过和水平高透过测量均较小(平均值±标准差:335.7±139.5µm、376.7±150.6µm、375.1±150.0µm,两者比较均 P<0.001)。
所提出的高透过测量是评估 FTMH 大小的一种可行且可重复的替代方法,并可为在本文所示的横断面谱域光相干断层扫描扫描或使用面向面投影在谱域光相干断层扫描容积数据集上自动测量 FTMH 提供基础。