Maffeis Giulia, Ferocino Edoardo, Mora Alberto Dalla, Pifferi Antonio, Cubeddu Rinaldo, Taroni Paola
Politecnico di Milano, Department of Physics, Piazza Leonardo da Vinci, 32, 20133, Milan, Italy.
CNR-Istituto di Fotonica e Nanotecnologie, Piazza Leonardo da Vinci, 32, 20133, Milan, Italy.
Biomed Opt Express. 2021 Jan 27;12(2):1105-1122. doi: 10.1364/BOE.412210. eCollection 2021 Feb 1.
A recent upgrade of the time domain multi-wavelength optical mammograph developed by Politecnico di Milano achieved good performance in laboratory tests [Biomed. Opt. Express9, 755 (2018).10.1364/BOE.9.000755]. However, it proved unsatisfactory when measurements were finally performed. That led to a further upgrade, including the replacement of the time-to-digital converter with a new model, and the related set-up changes. The new instrument version offers improved laboratory performance (as assessed through established protocols: BIP and MEDPHOT) and good performance (extension of the scanned breast area, repeatability, consistency of estimated tissue composition with physiology). Besides introducing the new set-up and detailing its laboratory and performance, we highlight the importance of systematic testing before entering clinical trials.