三种电子根尖定位仪准确性的实验室研究:嵌入介质的影响和电子根尖界限的放射评估。

A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit.

机构信息

Department of Endodontics, Life Sciences School, Pontifícia Universidade Católica do Paraná, Curitiba, Brazil.

Department of Periodontics and Endodontics, University at Buffalo, New York, NY, USA.

出版信息

Int Endod J. 2021 Jul;54(7):1200-1206. doi: 10.1111/iej.13507. Epub 2021 Mar 30.

Abstract

AIM

To evaluate several factors that might interfere with the use of electronic root canal length measurement devices (ERCLMDs) in a laboratory setting, including two different embedding media (alginate and electroconductive gel), three different types of devices and the radiographic view on the assessment of the electronic readings.

METHODOLOGY

Thirty single-rooted extracted human mandibular premolars were selected. After access and canal pre-flaring, a size 10 K-file was inserted in the canal up to the major apical foramen under magnification (×10), and this length was recorded as the actual length (AL) of the canal. Teeth were mounted on a specific endodontic training kit platform (ProTrain). Two different embedding media (electroconductive gel and alginate) and three different ERCLMDs: Apex ID (SybronEndo, Glendora, CA, USA), CanalPro Apex Locator (Coltene-Endo, Cuyahoga Falls, OH, USA) and the Root ZX II (J. Morita, Tokyo, Japan), were tested. Following the measurement at the 0.5 mark of each ERCLMD using alginate, two periapical radiographs (bucco-lingual and proximal views) were taken. The difference between the electronic readings and the AL, as well as the distance from the file tip to the apex of the roots on the radiographs, was calculated. Data were analysed statistically (two-way anova, Tukey's and chi-squared tests) at 5% significance level.

RESULTS

The incidence of over-extended readings was significantly greater (P < 0.05) when using the ProTrain electroconductive gel, except for Root ZX II at the 0.5 mark. Root ZX, CanalPro and Apex ID had similar accuracy when the 0.0/APEX mark was used to locate the foramen. For the radiographic method, no difference was found between the devices or radiographic views.

CONCLUSION

Compared with alginate, the conductive gel of the ProTrain kit negatively affected most of the electronic readings. The three evaluated ERCLMDs had similar accuracy in locating the apical foramen when using the mark 0.0/APEX and alginate as the embedding media. Although the ProTrain platform allows radiographs to be exposed in both bucco-lingual and proximal views, no difference was found between the views when evaluating the apical limit provided by the 0.5 mark in mandibular premolars embedded in alginate.

摘要

目的

评估在实验室环境中使用电子根管长度测量设备(ERCLMD)时可能干扰的几个因素,包括两种不同的嵌入介质(藻酸盐和导电凝胶)、三种不同类型的设备以及评估电子读数时的射线照相视图。

方法

选择 30 颗下颌前磨牙的单根牙。在进入和根管预扩孔后,将尺寸为 10 的 K 文件插入根管中,直至在放大倍数(×10)下的主要根尖孔,该长度记录为根管的实际长度(AL)。牙齿安装在特定的根管训练套件平台(ProTrain)上。测试了两种不同的嵌入介质(导电凝胶和藻酸盐)和三种不同的 ERCLMD:Apex ID(SybronEndo,Glendora,CA,USA)、CanalPro Apex Locator(Coltene-Endo,Cuyahoga Falls,OH,USA)和 Root ZX II(J. Morita,Tokyo,Japan)。在使用藻酸盐测量每个 ERCLMD 的 0.5 标记后,拍摄了两张根尖射线照相(颊舌向和近中向视图)。计算电子读数与 AL 之间的差值,以及文件尖端到根尖的距离。数据在 5%显着性水平下进行统计学分析(双向方差分析、Tukey 检验和卡方检验)。

结果

当使用 ProTrain 导电凝胶时,除了 Root ZX II 外,在 0.5 标记处过度延伸读数的发生率显着更高(P<0.05)。当使用 0.0/APEX 标记定位孔时,Root ZX、CanalPro 和 Apex ID 的准确性相似。对于射线照相方法,在设备或射线照相视图之间没有发现差异。

结论

与藻酸盐相比,ProTrain 试剂盒的导电凝胶对大多数电子读数产生了负面影响。当使用标记 0.0/APEX 和藻酸盐作为嵌入介质时,三种评估的 ERCLMD 在定位根尖孔方面具有相似的准确性。虽然 ProTrain 平台允许在颊舌向和近中向视图中暴露射线照相,但在评估下颌前磨牙中藻酸盐嵌入的 0.5 标记提供的根尖限制时,两种视图之间没有发现差异。