一种新型根尖定位仪的准确性:一项体外研究。

Accuracy of a new apex locator: an in vitro study.

作者信息

Kaufman A Y, Keila S, Yoshpe M

机构信息

Department of Endodontology, The Maurice and Gabriela Goldschleger School of Dental Medicine, Tel Aviv University, Tel Aviv, Israel.

出版信息

Int Endod J. 2002 Feb;35(2):186-92. doi: 10.1046/j.1365-2591.2002.00468.x.

Abstract

AIM

The purpose of this study was to test in an in vitro model the accuracy of a Bingo 1020 electronic apex locator, to compare the results to those of a well known apex locator, Root ZX, as well as to those of the radiographic method of tooth length determination.

METHODOLOGY

A total of 120 extracted teeth, preserved in Thymol solution and kept refrigerated, was used for the study. The experiment was performed on single-rooted teeth and on one-root canal, chosen randomly, in multirooted teeth. The teeth were randomly divided into 12 groups of 10 teeth each. After access preparation, the actual length (AL) was measured. The teeth were embedded in an alginate model specially developed for testing apex locators. Electronic tooth length measurements (EL) were carried out prior to root canal preparation using the two electronic apex locators (EAL) - Root ZX and Bingo 1020; three measurements were taken and an average computed. After the third measurement, the file was left in the root canal and a periapical radiograph was taken. The radiographic length (RL) was recorded by measuring the file length from the coronal reference point to the tip of the file. Each root canal was then prepared to a no. 40 K-file diameter using a standardized technique; saline was used for irrigation. Upon completion of the root canal preparation, EL measurements were taken by each EAL in dry conditions and with different irrigation solutions. Each measurement was repeated three times. The RL was recorded according to the last EL measurement. Results were subjected to statistical analysis.

RESULTS

In all parameters tested, a significant statistical difference was found between Bingo 1020 and the Root ZX. Measurements obtained using the Bingo 1020 were consistently closer to the AL (0.08 mm) than those obtained using the Root ZX. Both EALs measured the tooth length with great accuracy and a positive correlation of 0.76 (P = 0.00) existed between the two devices. No significant difference was found between the two apex locators when measurements were taken with the different irrigants (P = 0.34) and the content of the root canal did not affect the accuracy of the measurements. Lengths obtained by calculations from the radiographs were longer than the AL as well as the length obtained by both EALs (P = 0.00).

CONCLUSIONS

The Bingo 1020 proved to be as reliable as Root ZX and was user friendly. Under the experimental conditions, electronic measurements were more reliable than radiographs in the process of root length determination.

摘要

目的

本研究旨在通过体外模型测试Bingo 1020电子根尖定位仪的准确性,将结果与知名根尖定位仪Root ZX以及影像学确定牙齿长度的方法进行比较。

方法

本研究共使用120颗保存在百里酚溶液中并冷藏的离体牙。实验在单根牙和多根牙中随机选择的单根管上进行。牙齿被随机分为12组,每组10颗。开髓后,测量实际长度(AL)。将牙齿嵌入专门为测试根尖定位仪开发的藻酸盐模型中。在根管预备前,使用两种电子根尖定位仪(EAL)——Root ZX和Bingo 1020进行电子牙齿长度测量(EL);进行三次测量并计算平均值。第三次测量后,将锉留在根管内并拍摄根尖片。通过测量从冠部参考点到锉尖的长度记录影像学长度(RL)。然后使用标准化技术将每个根管预备到40号K锉直径;使用生理盐水冲洗。根管预备完成后,在干燥条件下和使用不同冲洗液的情况下,每种EAL进行EL测量。每次测量重复三次。根据最后一次EL测量记录RL。结果进行统计分析。

结果

在所有测试参数中,Bingo 1020和Root ZX之间存在显著统计学差异。使用Bingo 1020获得的测量值始终比使用Root ZX获得的测量值更接近AL(0.08毫米)。两种EAL测量牙齿长度的准确性都很高,两种设备之间存在正相关系数为0.76(P = 0.00)。当使用不同冲洗液进行测量时,两种根尖定位仪之间未发现显著差异(P = 0.34),根管内容物不影响测量的准确性。通过影像学计算获得的长度比AL以及两种EAL获得的长度都长(P = 0.00)。

结论

Bingo 1020被证明与Root ZX一样可靠且使用方便。在实验条件下,在确定牙根长度的过程中,电子测量比影像学测量更可靠。