Suppr超能文献

千焦耳 A500 二氧化碳吸收器大量泄漏并未抑制机械通气但手动皮囊通气:病例报告并实验再现。

Massive Leak in CO2 Absorber of Perseus A500 Did Not Inhibit Mechanical Ventilation but Manual Bag Ventilation: A Case Report With Experimental Reproduction.

机构信息

From the Departments of Anesthesiology.

Clinical Engineering, Kyorin University School of Medicine, Tokyo, Japan.

出版信息

A A Pract. 2021 Mar 19;15(3):e01425. doi: 10.1213/XAA.0000000000001425.

Abstract

Massive leaks in the anesthesia circuit may cause intraoperative hypoventilation and awareness; we experienced this with a disposable CO2 absorber in Perseus A500, which uses turbine ventilation to create positive-pressure ventilation. Consequently, manual ventilation was rendered impossible. During prolonged surgeries, CO2 absorbers may be replaced by a new one. In our case, the replacement had an occult leak. Absorbers should be checked before the exchange, and the econometer or reservoir bag's filling state should be monitored. Anesthesia providers should know an anesthesia machine's dynamics and breathing system to provide appropriate management of such a leak.

摘要

大量的麻醉回路泄漏可能导致术中通气不足和知晓;我们在使用涡轮通风机产生正压通气的 Perseus A500 一次性二氧化碳吸收器时遇到了这种情况。因此,手动通气变得不可能。在长时间的手术中,二氧化碳吸收器可能会被更换为新的。在我们的情况下,更换的吸收器有一个隐蔽的泄漏。更换前应检查吸收器,并且应监测经济计量器或储气囊的填充状态。麻醉提供者应了解麻醉机的动力学和呼吸系统,以对这种泄漏进行适当的管理。