Anode heel effect: Does it impact image quality in digital radiography? A systematic literature review.

机构信息

Department of Radiology & Nuclear Medicine, Hospital of Southwest Jutland, University Hospital of Southern Denmark, Finsensgade 35, 6700 Esbjerg, Denmark.

UCL University College, Niels Bohrs Allé 1, 5230 Odense M, Denmark.

出版信息

Radiography (Lond). 2021 Aug;27(3):976-981. doi: 10.1016/j.radi.2021.02.014. Epub 2021 Mar 17.

Abstract

INTRODUCTION

The anode heel effect can be used to optimize image quality and/or patient dose in digital radiography (DR). In film-screen radiography, the effect can equalize optical density in regions of varying attenuation. Clinical experience suggests that the implementation of DR has led to less awareness of anode orientation. Post-processing is assumed to compensate, but may also alter image impression and potentially obscure image details. Published evidence was examined for the influence of the anode heel effect on image quality in DR.

METHOD

A systematic literature search was carried out using PubMed, Embase, and Web of Science databases. Title and abstracts were screened blinded by three authors, according to in-/exclusion criteria, followed by full-text analysis for final inclusion. Studies where technical and/or visual image quality were reported, was included. All studies were analyzed and assigned quality scores, according to relevant questions. The authors devised a scoring system based on reported information pertaining to reproducibility, interpretation, and generalizability of the methods and conclusions.

RESULTS

Five studies were included of heterogeneous design, each with methodological shortcomings. Only a few anatomical areas were covered. Very few patients were examined, and in no studies were images evaluated by radiologists or reporting radiographers. Relevant information such as post-processing, image quality criteria and analysis was insufficient in most studies, making reproduction difficult. Results were contradictory, especially concerning technical vs visual image quality.

CONCLUSION

Limited published evidence was found quantifying the influence of the anode heel effect on image quality using DR technology. More methodologically, robust studies are needed. The published evidence neither proves nor disproves the impact of the heel effect on image quality in DR.

IMPLICATIONS FOR PRACTICE

Based on a systematic review, no firm recommendations for anode orientation relating to image quality in DR can be provided.

摘要

简介

阳极足跟效应可用于优化数字射线照相(DR)中的图像质量和/或患者剂量。在屏片射线照相中,该效应可以使不同衰减区域的光学密度均匀化。临床经验表明,DR 的实施导致人们对阳极方向的认识降低。人们认为后处理可以进行补偿,但也可能改变图像印象并潜在地掩盖图像细节。本文旨在检查阳极足跟效应对 DR 中图像质量的影响的相关证据。

方法

使用 PubMed、Embase 和 Web of Science 数据库进行了系统的文献检索。三位作者根据纳入/排除标准对标题和摘要进行了盲法筛选,然后进行全文分析以最终纳入。纳入了报告技术和/或视觉图像质量的研究。根据相关问题,对所有研究进行了分析并进行了质量评分。作者根据报告的信息制定了一个评分系统,这些信息涉及方法和结论的可重复性、解释和可推广性。

结果

纳入了五项设计各异的研究,每项研究都存在方法学上的缺陷。仅涵盖了少数几个解剖区域。检查的患者很少,并且在没有研究中由放射科医生或报告放射技师评估图像。在大多数研究中,后处理、图像质量标准和分析等相关信息不足,使得结果难以重现。结果相互矛盾,尤其是在技术和视觉图像质量方面。

结论

使用 DR 技术量化阳极足跟效应对图像质量的影响的证据有限。需要更多的方法学上稳健的研究。已发表的证据既不能证明也不能反驳足跟效应对 DR 中图像质量的影响。

实践意义

基于系统评价,无法针对 DR 中与图像质量相关的阳极方位提供明确的建议。