Shrive N G, Lam T C, Damson E, Frank C B
Department of Civil Engineering, University of Calgary, Alberta, Canada.
J Biomech Eng. 1988 May;110(2):104-9. doi: 10.1115/1.3108413.
There appears to be no generally accepted method of measuring in-situ the cross-sectional area of connective tissues, particularly small ones, before mechanical testing. An instrument has therefore been devised to measure the cross-sectional area of one such tissue, the rabbit medial collateral ligament, directly and nondestructively. However, the methodology is general and could be applied to other tissues with appropriate changes in detail. The concept employed in the instrument is to measure the thickness of the tissue as a function of position along the width of the tissue. The plot obtained of thickness versus width position is integrated to provide the cross-sectional area. This area is accurate to within 5 percent, depending mainly on alignment of the instrument and pre-load of the ligament. Results on the mid-substance of the rabbit medial collateral ligaments are repeatable and reproducible. Values of maximum width and thickness are less variable than those obtained with a vernier caliper. The measured area is considerably less than that estimated assuming rectangular cross-section and slightly less than that estimated on the assumption of elliptical cross-section.
在进行力学测试之前,似乎没有一种普遍接受的方法可以原位测量结缔组织的横截面积,尤其是小尺寸的结缔组织。因此,设计了一种仪器,用于直接且无损地测量一种这样的组织——兔内侧副韧带的横截面积。然而,该方法具有通用性,只需在细节上做适当改变,就可应用于其他组织。该仪器采用的概念是测量组织厚度随沿组织宽度方向位置的变化。将得到的厚度与宽度位置的关系图进行积分,以提供横截面积。该面积的测量精度在5%以内,主要取决于仪器的校准和韧带的预加载。兔内侧副韧带中部的测量结果具有可重复性和再现性。最大宽度和厚度的值比用游标卡尺测得的更稳定。测得的面积远小于假设为矩形横截面时估计的面积,略小于假设为椭圆形横截面时估计的面积。