Suppr超能文献

两种不同的全髋关节骨水泥型置换术的影像学表现。一项生存分析。

Radiographic performance of two different total hip cemented arthroplasties. A survivorship analysis.

作者信息

Sarmiento A, Natarajan V, Gruen T A, McMahon M

机构信息

Department of Orthopaedics, UCLA School of Medicine.

出版信息

Orthop Clin North Am. 1988 Jul;19(3):505-15.

PMID:3380529
Abstract

Detailed serial radiographic analysis of the femoral and acetabular components of 367 Charnley (cobalt-chrome alloy) and 383 STH (titanium alloy) primary cemented total hip prostheses was conducted. The follow-up of the Charnley prosthesis ranged from 1 to 16 years (mean follow-up, 9 years), and for the STH from 1 to 11 years (mean follow-up, 80 months). Both single period and survivorship analyses were done to evaluate the radiographic performance. The analysis revealed that the curved STH prostheses' overall radiographic performance was less satisfactory than the Charnley prostheses. However, the STH with a straight stem had significantly better radiographic performance compared with the Charnley prostheses in all parameters except in the incidence of bone-cement radiolucent lines.

摘要

对367例Charnley型(钴铬合金)和383例STH型(钛合金)初次骨水泥固定全髋关节假体的股骨和髋臼组件进行了详细的系列影像学分析。Charnley假体的随访时间为1至16年(平均随访9年),STH假体的随访时间为1至11年(平均随访80个月)。进行了单期分析和生存率分析以评估影像学表现。分析显示,弯曲的STH假体的总体影像学表现不如Charnley假体。然而,除骨水泥透亮线发生率外,直柄STH假体在所有参数上的影像学表现均显著优于Charnley假体。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验