与Lubinus假体相比,ITH(可能是某种特定类型,此处未明确全称)假体的柄部松动更为常见。对142例患者进行了为期5年的临床和影像学随访。

Stem loosening more common with ITH than Lubinus prosthesis. A 5-year clinical and radiographic follow-up of 142 patients.

作者信息

Jacobsson S A, Ivarsson I, Djerf K, Wahlström O

机构信息

Department of Orthopedics, University Hospital of Linköping, University Hospital, Sweden.

出版信息

Acta Orthop Scand. 1995 Oct;66(5):425-31. doi: 10.3109/17453679508995579.

Abstract

Out of a consecutive series of 169 cemented total hip arthroplasties, 77 Lubinus SP 2 and 65 ITH prostheses, were followed-up after 5 years with clinical and radiographic assessments. Radiolucencies were commoner around the ITH titanium stem than around the Lubinus cobalt-chrome stem. 6 of the ITH stems were considered as definitely loose, as compared to none of the Lubinus stems. Hitherto, 5 patients in the ITH group have had revisions and 2 need revision because of loosening. The poorer performance of the cemented ITH stem may be due to an unfavorable combination of design and material properties.

摘要

在连续的169例骨水泥型全髋关节置换术中,对77例Lubinus SP 2假体和65例ITH假体进行了5年的随访,包括临床和影像学评估。ITH钛柄周围的透亮线比Lubinus钴铬柄周围更常见。ITH柄中有6个被确定为松动,而Lubinus柄无一松动。迄今为止,ITH组有5例患者进行了翻修,2例因松动需要翻修。骨水泥型ITH柄性能较差可能是由于设计和材料特性的不利组合。