Conventional stylet-driven leads with extendable helix can be implanted successfully for left bundle branch area pacing (LBBAP) with a low acute complication rate. We report two cases in which lead repositioning after a first unsuccessful attempt to LBBAP was associated with fracture of the helix rotating mechanism and failure to fully extract the pacing lead.