González-Sanchis A, Brualla-González L, Fuster-Diana C, Gordo-Partearroyo J C, Piñeiro-Vidal T, García-Hernandez T, López-Torrecilla J L
Department of Radiation Oncology, ERESA, Hospital General Universitario de Valencia (CHGUV), Avenida Tres Cruces, No. 2, 46014, Valencia, Spain.
Department of Radiophysics, ERESA, Hospital General Universitario de Valencia (CHGUV), Valencia, Spain.
Clin Transl Oncol. 2021 Oct;23(10):2120-2126. doi: 10.1007/s12094-021-02617-6. Epub 2021 Apr 11.
Hypofractionated radiation therapy for breast cancer requires highly precise delivery through the use of image-guided radiotherapy (IGRT). Surface-guided radiation therapy (SGRT) is being increasingly used for patient positioning in breast radiotherapy. We aimed to assess the role of SGRT for verification of breast radiotherapy and the tumour bed.
Prospective study of 252 patients with early stage breast cancer. A total of 1170 determinations of daily positioning were performed. Breast surface positioning was determined with SGRT (AlignRT) and correlated with the surgical clips in the tumour bed, verified by IGRT (ExacTrac).
SGRT improved surface matching by a mean of 5.3 points compared to conventional skin markers (98.0 vs. 92.7), a statistically significant difference (p < 0.01, Wilcoxon Test). For surface matching values > 95%, ≥ 3 clips coincided in 99.7% of the determinations and all markers coincided in 92.5%. For surface matching rates > 90%, the location of ≥ 3 clips coincided in 99.55% of determinations.
SGRT improves patient positioning accuracy compared to skin markers. Optimal breast SGRT can accurately verify the localisation of the tumour bed, ensuring matching with ≥ 3 surgical clips. SGRT can eliminate unwanted radiation from IGRT verification systems.
乳腺癌的大分割放疗需要通过图像引导放疗(IGRT)实现高度精确的照射。表面引导放疗(SGRT)在乳腺癌放疗的患者定位中应用越来越广泛。我们旨在评估SGRT在乳腺癌放疗及瘤床验证中的作用。
对252例早期乳腺癌患者进行前瞻性研究。共进行了1170次每日定位测定。采用SGRT(AlignRT)确定乳房表面定位,并与瘤床中的手术夹进行关联,通过IGRT(ExacTrac)进行验证。
与传统皮肤标记相比,SGRT使表面匹配平均提高了5.3分(98.0对92.7),差异具有统计学意义(p < 0.01,Wilcoxon检验)。对于表面匹配值> 95%,在99.7%的测定中≥3个夹子重合,所有标记物在92.5%的测定中重合。对于表面匹配率> 90%,在99.55%的测定中≥3个夹子的位置重合。
与皮肤标记相比,SGRT提高了患者定位的准确性。最佳的乳房SGRT能够准确验证瘤床的定位,确保与≥3个手术夹匹配。SGRT可以消除IGRT验证系统中不必要的辐射。