Donnelly J R, Sovocool G W, Mitchum R K
Lockheed-EMSCO, Las Vegas, NV 89119.
J Assoc Off Anal Chem. 1988 Mar-Apr;71(2):434-9.
Mass and intensity calibration of gas chromatograph/mass spectrometer (GC/MS) responses is an important quality assurance issue for chemical analysis. Ion abundance calibration with decafluorotriphenylphosphine (DFTPP) was applied in 1975 to standardize quadrupole spectra to resemble the ion abundances that were obtainable from magnetic sector mass spectrometers. Modern quadrupole mass spectrometers provide significantly greater high-mass sensitivity than allowed under the 1975 study. Thus, those recommendations were reevaluated with 2 approaches. First, an interlaboratory study was conducted using 15 different gas chromatography/mass spectrometry (GC/MS) systems. Second, the U.S. Environmental Protection Agency Contract Laboratory Program (EPA-CLP) quality assurance data base was searched and over 6500 DFTPP tune results were plotted and evaluated. Based on these approaches, updated ion abundance criteria recommendations have been developed, which contemporary instruments can meet, and which meet data quality objectives regarding identification and quantitative analysis of analytes.
气相色谱/质谱仪(GC/MS)响应的质量和强度校准是化学分析中一个重要的质量保证问题。1975年采用十氟三苯基膦(DFTPP)进行离子丰度校准,以使四极杆质谱标准化,使其离子丰度类似于磁扇形质谱仪可获得的离子丰度。现代四极杆质谱仪提供的高质量灵敏度比1975年的研究允许的灵敏度高得多。因此,采用两种方法对这些建议进行了重新评估。首先,使用15种不同的气相色谱/质谱(GC/MS)系统进行了一项实验室间研究。其次,搜索了美国环境保护局合同实验室计划(EPA-CLP)质量保证数据库,并绘制和评估了6500多个DFTPP调谐结果。基于这些方法,制定了更新的离子丰度标准建议,当代仪器可以满足这些建议,并且这些建议符合关于分析物鉴定和定量分析的数据质量目标。