Suppr超能文献

应用 3D Navigo 系统靶向前列腺活检检测前列腺癌:一项前瞻性研究。

Prostate cancer detection by targeted prostate biopsy using the 3D Navigo system: a prospective study.

机构信息

Urology Department, University Hospital of Angers, 4 rue Larrey, 49933, Angers Cedex 9, France.

Radiology Department, University Hospital of Angers, Angers, France.

出版信息

Abdom Radiol (NY). 2021 Sep;46(9):4381-4387. doi: 10.1007/s00261-021-03078-9. Epub 2021 Apr 15.

Abstract

PURPOSE

The 3D Navigo™ system is a magnetic resonance imaging (MRI) and transrectal ultrasound (TRUS) fusion device for prostate targeted biopsies (TB). Our aim was to evaluate the clinically significant prostate cancer (CSC) detection rate of TB using the 3D Navigo™ system.

METHODS

Patients who underwent TB with the 3D Navigo™ system in our center between June 2014 and May 2018 were prospectively included, excluding those who have previously received treatment for prostate cancer. A 3-Tesla MRI imaging was performed before biopsies; findings were reported according to the Prostate Imaging Reporting and Data System version 2 (PIRADS). CSC was defined by an ISUP score ≥ 2.

RESULTS

304 patients underwent TB. Median age was 66 years (51-84). Median PSA was 7.75 ng/ml (0.6-70.0). Median prostate volume was 45.0 ml (15.9-221.7). PCa and CSC were found in 70.4% (214/304) and 47.7% (145/304) of the patients, respectively. The proportion of patients diagnosed with CSC among those with PCa was 67.8% (145/214). There was a significant risk of having a CSC in case of PIRADS score ≥ 4 and 5 (OR 5.0, 95% CI [2.7-9.2], P < 0.001; OR 3.2, 95% CI [1.8-5.5], P < 0.001). PIRADS score was an independent risk factor of having a CSC (OR 4.19, 95% CI [2.49-7.05], P < 0.001). There was no significant difference between pathological outcomes of TB and RP in paired analysis (P = 0.892). There was a correlation between TB and RP specimens for PCa detection (r = 0.60, P < 0.001).

CONCLUSION

Detecting CSC with MRI-TRUS fusion targeted biopsies using the 3D Navigo™ system is feasible and safe. We found a positive correlation between TB and RP for ISUP scores.

摘要

目的

3D Navigo™系统是一种用于前列腺靶向活检(TB)的磁共振成像(MRI)和经直肠超声(TRUS)融合设备。我们的目的是评估使用 3D Navigo™系统进行 TB 的临床显著前列腺癌(CSC)检出率。

方法

本研究前瞻性纳入了 2014 年 6 月至 2018 年 5 月在我院接受 3D Navigo™系统 TB 的患者,排除了之前接受过前列腺癌治疗的患者。在活检前进行 3-Tesla MRI 成像;根据前列腺成像报告和数据系统第 2 版(PIRADS)报告结果。CSC 定义为 ISUP 评分≥2。

结果

共 304 例患者接受了 TB。中位年龄为 66 岁(51-84 岁)。中位 PSA 为 7.75ng/ml(0.6-70.0)。中位前列腺体积为 45.0ml(15.9-221.7)。PCa 和 CSC 在 70.4%(214/304)和 47.7%(145/304)的患者中分别发现。在患有 PCa 的患者中,诊断为 CSC 的患者比例为 67.8%(145/214)。PIRADS 评分≥4 和 5 时,CSC 的风险显著增加(OR 5.0,95%CI [2.7-9.2],P<0.001;OR 3.2,95%CI [1.8-5.5],P<0.001)。PIRADS 评分是 CSC 的独立危险因素(OR 4.19,95%CI [2.49-7.05],P<0.001)。TB 和 RP 在配对分析中的病理结果无显著差异(P=0.892)。TB 和 RP 标本在 PCa 检测方面存在相关性(r=0.60,P<0.001)。

结论

使用 3D Navigo™系统进行 MRI-TRUS 融合靶向活检检测 CSC 是可行且安全的。我们发现 TB 和 RP 在 ISUP 评分方面存在正相关。