Instituto de Física 'Gleb Wataghin', Universidade Estadual de Campinas, 13083-859, Campinas, Brazil.
Facultat de Física (FQA and ICC), Universitat de Barcelona, Diagonal 645, ES-08028 Barcelona, Catalonia, Spain.
Phys Med Biol. 2021 May 25;66(11). doi: 10.1088/1361-6560/abf859.
Monte Carlo (MC) simulations are employed extensively in breast dosimetry studies. In the energy interval of interest in mammography energy deposition is predominantly caused by the photoelectric effect, and the corresponding cross sections used by the MC codes to model this interaction process have a direct influence on the simulation results. The present work compares two photoelectric cross section databases in order to estimate the systematic uncertainty, related to breast dosimetry, introduced by the choice of cross sections for photoabsorption. The databases with and without the so-called normalization screening correction are denoted as 'renormalized' or 'un-normalized', respectively. The simulations were performed with the PENELOPE/penEasy code system, for a geometry resembling a mammography examination. The mean glandular dose (MGD), incident air kerma (), normalized glandular dose (DgN) and glandular depth-dose (GDD()) were scored, for homogeneous breast phantoms, using both databases. The AAPM Report TG-195 case 3 was replicated, and the results were included. Moreover, cases with heterogeneous and anthropomorphic breast phantoms were also addressed. The results simulated with the un-normalized cross sections are in better overall agreement with the TG-195 data than those from the renormalized cross sections; for MGD the largest discrepancies are 0.13(6)% and 0.74(5)%, respectively. The MGD,and DgN values simulated with the two databases show differences that diminish from approximately 10%/3%/6.8% at 8.25 keV down to 1.5%/1.7%/0.4% at 48.75 keV, respectively. For polyenergetic spectra, deviations up to 2.5% were observed. The disagreement between the GDDs simulated with the analyzed databases increases with depth, ranging from -1% near the breast entrance to 4% near the bottom. Thus, the choice of photoelectric cross section database affects the MC simulation results of breast dosimetry and adds a non-negligible systematic uncertainty to the dosimetric quantities used in mammography.
蒙特卡罗 (MC) 模拟在乳房剂量学研究中得到了广泛应用。在乳腺摄影能量范围内,能量沉积主要由光电效应引起,MC 代码用于模拟这一相互作用过程的相应截面直接影响模拟结果。本工作比较了两种光电截面数据库,以估计由于光吸收截面选择而导致的与乳房剂量学相关的系统不确定性。具有和不具有所谓的归一化筛选校正的数据库分别表示为“再归一化”或“非归一化”。模拟使用 PENELOPE/penEasy 代码系统进行,几何形状类似于乳腺摄影检查。对于均匀乳房体模,使用这两个数据库对平均腺体剂量 (MGD)、入射空气比释动能 ()、归一化腺体剂量 (DgN) 和腺体深度剂量 (GDD()) 进行评分。复制了 AAPM 报告 TG-195 案例 3,并包含了结果。此外,还处理了具有异质和人体模型乳房体模的情况。使用非归一化截面模拟的结果与 TG-195 数据的总体一致性更好,与再归一化截面的结果相比,最大差异分别为 0.13(6)%和 0.74(5)%。使用这两个数据库模拟的 MGD 和 DgN 值的差异从 8.25 keV 时的约 10%/3%/6.8%逐渐减小到 48.75 keV 时的 1.5%/1.7%/0.4%。对于多能谱,观察到高达 2.5%的偏差。分析数据库模拟的 GDD 之间的差异随深度增加而增加,从乳房入口附近的-1%到底部附近的 4%。因此,光电截面数据库的选择会影响乳房剂量学的 MC 模拟结果,并为乳腺摄影中使用的剂量学量添加不可忽略的系统不确定性。