Allman T, Holland G A, Lenkinski R E, Charles H C
Department of Radiology, University of Pennsylvania, Philadelphia 19104.
Magn Reson Med. 1988 May;7(1):88-94. doi: 10.1002/mrm.1910070110.
It is a requirement of many localized NMR spectroscopy experiments, such as DRESS, that collection of the free induction decay be delayed a significant time after the middle of the last pulse. This leads to distortion of the spectra and seriously impedes the retrieval of quantitative information from these spectra. A technique is described in which a correction function is derived from the convoluted spectrum and is used to produce a spectrum with much reduced distortion from which quantitative information may easily be obtained by conventional techniques.
许多局部核磁共振光谱实验(如DRESS)都要求在最后一个脉冲的中点之后延迟相当长的时间来采集自由感应衰减信号。这会导致光谱失真,并严重阻碍从这些光谱中获取定量信息。本文描述了一种技术,其中从卷积光谱中导出校正函数,并用于生成失真大大降低的光谱,通过传统技术可以很容易地从该光谱中获得定量信息。