Hovgesen Caroline Harder, Wilhjelm Jens E, Vilmann Peter, Kalaitzakis Evangelos
Biomedical Engineering, Department of Health Technology, Technical University of Denmark, Kongens Lyngby, Denmark.
Gastrounit, Herlev University Hospital, Herlev, Denmark.
J Ultrasound Med. 2022 Feb;41(2):311-325. doi: 10.1002/jum.15713. Epub 2021 Apr 18.
Optimal visualization of needles in clinical ultrasound imaging is important and challenging, especially at steep angles. Improvement of visualization has been attempted with various techniques, for example, coatings and dimples. This systematic review summarizes enhancement techniques and identifies superior echogenic surface enhancements. Twenty-four papers were identified providing visibility measures for 33 different echogenic needles. These were grouped according to surface characteristics and ranked. Echogenic needles ranked higher than standard needles especially at steeper angles. Among the echogenic needles, coated needles were seemingly better visualized "in vivo" than noncoated needles, despite heterogeneity in study conditions. No unambiguous comparison revealed which needle was best visualized.
在临床超声成像中,实现针的最佳可视化既重要又具有挑战性,尤其是在大角度时。人们已尝试用各种技术来改善可视化效果,例如涂层和凹坑。本系统评价总结了增强技术,并确定了卓越的回声增强表面。共识别出24篇论文,这些论文提供了33种不同回声针的可视化测量方法。根据表面特征对这些针进行了分组并排序。回声针的排名高于标准针,尤其是在较大角度时。在回声针中,尽管研究条件存在异质性,但涂层针在“体内”的可视化效果似乎优于未涂层针。没有明确的比较表明哪种针的可视化效果最佳。