Prehospital coagulation measurement by a portable blood analyzer in a helicopter emergency medical service (HEMS).

机构信息

Department of Anesthesiology, Amsterdam University Medical Center, Amsterdam, Netherlands; HEMS Life Liner 1, Amsterdam University Medical Center, Amsterdam, Netherlands.

Department of Anesthesiology, Amsterdam University Medical Center, Amsterdam, Netherlands.

出版信息

Am J Emerg Med. 2021 Aug;46:137-140. doi: 10.1016/j.ajem.2021.04.023. Epub 2021 Apr 19.

Abstract

In helicopter emergency medical services, HEMS, coagulopathy presents both in trauma (e.g. consumption of coagulation factors) and non-trauma cases (e.g. anticoagulant use). Therefore, in HEMS coagulation measurements appear promising and Prothrombin Time (PT) and derived INR are attractive variables herein. We tested the feasibility of prehospital PT/INR coagulation measurements in HEMS. This study was performed at the Dutch HEMS, using a portable blood analyzer (i-Stat®1, Abbott). PT/INR measurements were performed on (hemodiluted) author's blood, and both trauma- and non-trauma HEMS patients. Device-related benefits of the i-Stat PT/INR system were portability, speed and ease of handling. Limitations included a rather narrow operational temperature range (16-30 °C). PT/INR measurements (n = 15) were performed on hemodiluted blood, and both trauma and non-trauma patients. The PT/INR results confirmed effects of hemodilution and anticoagulation, however, most measurement results were in the normal INR-range (0.9-1.2). We conclude that prehospital PT/INR measurements, although with limitations, are feasible in HEMS operations.

摘要

在直升机紧急医疗服务(HEMS)中,凝血障碍既存在于创伤(例如凝血因子消耗)病例中,也存在于非创伤病例(例如抗凝剂使用)中。因此,在 HEMS 中,凝血测量似乎很有前景,凝血酶原时间(PT)和衍生的国际标准化比值(INR)是很有吸引力的变量。我们测试了在 HEMS 中进行院前 PT/INR 凝血测量的可行性。这项研究是在荷兰 HEMS 进行的,使用了便携式血液分析仪(i-Stat®1,雅培)。PT/INR 测量是在(血液稀释的)作者的血液以及创伤和非创伤 HEMS 患者中进行的。i-Stat PT/INR 系统的设备相关优势包括便携性、速度和易于操作。局限性包括相对较窄的操作温度范围(16-30°C)。PT/INR 测量(n = 15)是在血液稀释的血液和创伤和非创伤患者中进行的。PT/INR 结果证实了血液稀释和抗凝的影响,但大多数测量结果都在正常 INR 范围内(0.9-1.2)。我们得出结论,尽管存在局限性,但院前 PT/INR 测量在 HEMS 操作中是可行的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验