Agreement Between Retinoscopy, Autorefractometry and Subjective Refraction for Determining Refractive Errors in Congolese Children.

作者信息

Mukash Sabrina N, Kayembe David L, Mwanza Jean-Claude

机构信息

Department of Ophthalmology, School of Medicine, Kinshasa University Hospital, Kinshasa, Democratic Republic of Congo.

Department of Ophthalmology, School of Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC, USA.

出版信息

Clin Optom (Auckl). 2021 Apr 21;13:129-136. doi: 10.2147/OPTO.S303286. eCollection 2021.

Abstract

OBJECTIVE

To assess the agreement between retinoscopy and autorefractometry and between subjective refraction and both retinoscopy and autorefractometry in Congolese children.

PATIENTS AND METHODS

Fifty-four children (6-17 years old) were enrolled consecutively in this cross-sectional study. Refraction was evaluated before and after cycloplegia (1% cyclopentolate) with retinoscopy and autorefractometry. Readings were compared (paired -test) and agreement assessed with Bland-Altman plots. Subjective refraction was compared with the two methods to determine which one provides better reference estimates for subjective refraction.

RESULTS

Under cycloplegia, the spherical power was comparable between retinoscopy and autorefractometry (1.12 ± 1.37 D vs 1.22 ± 1.06D, P = 0.70), cylinder power was significantly more myopic on retinoscopy than autorefractometry (0.80 ± 1.10D vs -0.62 ± 0.66, P = 0.019), and SE was greater on autorefractometry than retinoscopy (0.91 ± 1.10D vs 0.72 ± 1.00D, P = 0.014). Retinoscopy and autorefractometry overestimated the power of spherical (P = 0.022 and 0.002, respectively) and cylindrical components (all P < 0.001). There was an agreement between retinoscopy and autorefractometry in measuring spherical (bias: 0.09 ± 0.16D; limit of agreement, LoA: -0.40 to 0.22D) and cylindrical power (bias: -0.18 ± 0.20D; LoA: -0.57 to 0.21D). Subjective refraction agreed with cycloplegic retinoscopy for determining SE power (bias: 0.11D; LoA: -0.51 to 0.73D).

CONCLUSION

Retinoscopy and autorefractometry can be used interchangeably in children for determining the power of spherical and cylindrical components. Cycloplegic retinoscopy is better than autorefractometry to obtain SE reference values for subjective refraction in children.

摘要

目的

评估刚果儿童视网膜检影法与自动验光仪之间以及主观验光与视网膜检影法和自动验光仪之间的一致性。

患者与方法

54名儿童(6至17岁)连续纳入本横断面研究。在使用1%环戊通散瞳前后,采用视网膜检影法和自动验光仪评估屈光情况。对读数进行比较(配对t检验),并用布兰德-奥特曼图评估一致性。将主观验光与这两种方法进行比较,以确定哪种方法能为主观验光提供更好的参考估计值。

结果

散瞳状态下,视网膜检影法与自动验光仪测得的球镜度数相当(1.12±1.37D对1.22±1.06D,P = 0.70),视网膜检影法测得的柱镜度数比自动验光仪显著更近视(0.80±1.10D对 -0.62±0.66,P = 0.019),自动验光仪测得的等效球镜度数比视网膜检影法更大(0.91±1.10D对0.72±1.00D,P = 0.014)。视网膜检影法和自动验光仪均高估了球镜(分别为P = 0.022和0.002)和柱镜成分(所有P < 0.001)的度数。视网膜检影法与自动验光仪在测量球镜度数(偏差:0.09±0.16D;一致性界限,LoA:-0.40至0.22D)和柱镜度数(偏差:-0.18±0.20D;LoA:-0.57至0.21D)方面具有一致性。主观验光与散瞳后视网膜检影法在确定等效球镜度数方面具有一致性(偏差:0.11D;LoA:-0.51至0.73D)。

结论

在儿童中,视网膜检影法和自动验光仪可互换使用以确定球镜和柱镜成分的度数。散瞳后视网膜检影法在获取儿童主观验光的等效球镜参考值方面优于自动验光仪。