Qian K X, Fei Q
Shanghai Second Medical University, China.
Clin Phys Physiol Meas. 1988 May;9(2):107-12. doi: 10.1088/0143-0815/9/2/002.
Non-pulsatile and pulsatile impeller blood pumps have been developed. The non-pulsatile pump was compared with the Shanghai and Sarns 7000 roller pumps for haemolysis. Similarly, the pulsatile pump was compared with a pneumatic diaphragm pump and the Polystan pulsatile pump. In each test, one impeller pump and the pump it was being compared with were connected to two identical circulatory systems with the same volume of fresh citrated porcine blood, delivered at the same pressure and flow for several hours. Every half hour the free haemoglobin was measured and the index of haemolysis (IH) calculated. The mean IH of the non-pulsatile impeller pump was about 18% of the Shanghai roller pump and 29% of the Sarns 7000 roller pump; the IH of the pulsatile impeller pump was 16% of the diaphragm pump and 8% of the Polystan pulsatile pump.
非搏动性和搏动性叶轮血泵已被研发出来。将非搏动性血泵与上海血泵和Sarns 7000滚压泵进行溶血比较。同样,将搏动性血泵与气动隔膜泵和Polystan搏动性血泵进行比较。在每次测试中,一台叶轮泵及其被比较的泵连接到两个相同的循环系统,系统中装有相同体积的新鲜枸橼酸化猪血,以相同的压力和流量输送数小时。每半小时测量一次游离血红蛋白,并计算溶血指数(IH)。非搏动性叶轮血泵的平均IH约为上海滚压泵的18%和Sarns 7000滚压泵的29%;搏动性叶轮血泵的IH为隔膜泵的16%和Polystan搏动性血泵的8%。