Lambrecht L, Vanholder R, Ringoir S
Department of Medicine, University Hospital, Ghent, Belgium.
Blood Purif. 1988;6(2):77-84. doi: 10.1159/000169487.
Plasma filtration characteristics of three hollow fiber plasma separation filters (Plasmaflo Hi-05, Extraplex BL 550 and CPS-10) were studied in a single-needle setting by means of the double headpump. Plasma exchange was carried out in 12 patients during a total number of 59 sessions. For each filter, a mean total ultrafiltration volume of +/- 3,000 ml was obtained over a period from 92 to 121 min. The lowest and highest obtained mean filtration flows were 26.7 +/- 2.5 and 36.6 +/- 1.7 ml/min for Extraplex BL 550 and CPS-10, respectively (p less than 0.01). The pre- and postplasmapheresis pressures, measured in the bubble trap chamber as an indirect estimation of transmembrane pressure, were lower for the Plasmaflo Hi-05 than for the two other filters under study; pressures remained unaltered during the session. Blood pressure showed a minor but significant decline during plasmapheresis with the Plasmaflo filter. A reduction after plasmapheresis by more than 40% of the immunoglobulins IgE, IgG, IgM and IgA, and of complement factors C3 and C4 was seen for each of the filters and no significant differences between the filters were observed. An additional study on 6 filters with constant blood flow and TMP showed minor differences in the transmembrane pressure necessary to obtain a given filtration volume per unit of time and similar sieving coefficients for immunoglobulins. This study demonstrates that with this single-needle technique a satisfying immunoglobulin extraction performance was obtained for each of the filter types studied; however, there existed minor but significant differences in the patient hemodynamic status according to the membrane used.
采用双头泵在单针模式下研究了三种中空纤维血浆分离滤器(Plasmaflo Hi-05、Extraplex BL 550和CPS-10)的血浆过滤特性。在12例患者中进行了总共59次血浆置换。对于每种滤器,在92至121分钟的时间段内获得的平均总超滤量为±3000 ml。Extraplex BL 550和CPS-10获得的最低和最高平均过滤流量分别为26.7±2.5和36.6±1.7 ml/分钟(p<0.01)。作为跨膜压力的间接估计,在气泡捕集室中测量的血浆置换前后压力,Plasmaflo Hi-05比其他两种研究中的滤器要低;在整个过程中压力保持不变。使用Plasmaflo滤器进行血浆置换时,血压出现轻微但显著的下降。每种滤器在血浆置换后免疫球蛋白IgE、IgG、IgM和IgA以及补体因子C3和C4均降低了40%以上,且未观察到滤器之间存在显著差异。另一项对6个滤器在恒定血流和跨膜压下的研究表明,在获得单位时间给定过滤体积所需的跨膜压力方面存在微小差异,而免疫球蛋白的筛分系数相似。本研究表明,采用这种单针技术,所研究的每种滤器类型都能获得令人满意的免疫球蛋白提取性能;然而,根据所使用的膜,患者的血流动力学状态存在微小但显著的差异。