Hughes Luke, McEwen James
Centre for Applied Performance Sciences, Faculty of Sport, Allied Health and Performance Sciences, St Mary's University, London, UK.
Department of Orthopaedics, Faculty of Medicine, University of British Columbia, Vancouver, Canada.
BMC Biomed Eng. 2021 May 8;3(1):8. doi: 10.1186/s42490-021-00053-9.
Development of automatic, pneumatic tourniquet technology and use of personalised tourniquet pressures has improved the safety and accuracy of surgical tourniquet systems. Personalisation of tourniquet pressure requires accurate measurement of limb occlusion pressure (LOP), which can be measured automatically through two different methods. The 'embedded LOP' method measures LOP using a dual-purpose tourniquet cuff acting as both patient sensor and pneumatic effector. The 'distal LOP' method measures LOP using a distal sensor applied to the patient's finger or toe of the operating limb, using photoplethysmography to detect volumetric changes in peripheral blood circulation. The distal LOP method has been used clinically for many years; the embedded LOP method was developed recently with several advantages over the distal LOP method. While both methods have clinically acceptable accuracy in comparison to LOP measured using the manual Doppler ultrasound method, these two automatic methods have not been directly compared. The purpose of this study is to investigate if the embedded and distal methods of LOP measurement have clinically acceptable agreement. The differences in pairs of LOP measurement in the upper and lower limbs of 81 healthy individuals were compared using modified Bland and Altman analysis. In surgery, it is common for cuff pressure to deviate from the pressure setpoint due to limb manipulation. Surgical tourniquet systems utilise a ± 15 mmHg pressure alarm window, whereby if the cuff pressure deviates from the pressure setpoint by > 15 mmHg, an audiovisual alarm is triggered. Therefore, if the difference (bias) ± SE, 95% CI of the bias and SD of differences ± SE in LOP measurement between the embedded and distal methods were all within ±15 mmHg, this would demonstrate that the two methods have clinically acceptable agreement.
LOP measurement using the embedded LOP method was - 0.81 ± 0.75 mmHg (bias ± standard error) lower than the distal LOP method. The 95% confidence interval of the bias was - 2.29 to 0.66 mmHg. The standard deviation of the differences ± standard error was 10.35 ± 0.49 mmHg. These results show that the embedded and distal methods of LOP measurement demonstrate clinically acceptable agreement.
The findings of this study demonstrate clinically acceptable agreement between the embedded and distal methods of LOP measurement. The findings support the use of the embedded LOP method of automatic LOP measurement using dual-purpose tourniquet cuffs to enable accurate, effective and simple prescription of personalised tourniquet cuff pressures in a clinical setting.
自动气压止血带技术的发展以及个性化止血带压力的应用提高了手术止血带系统的安全性和准确性。止血带压力的个性化需要准确测量肢体阻断压力(LOP),这可以通过两种不同的方法自动测量。“嵌入式LOP”方法使用兼作患者传感器和气动效应器的两用止血带袖带测量LOP。“远端LOP”方法使用应用于手术肢体患者手指或脚趾的远端传感器,通过光电容积描记法检测外周血液循环中的容积变化来测量LOP。远端LOP方法已在临床上使用多年;嵌入式LOP方法是最近开发的,比远端LOP方法有几个优点。虽然与使用手动多普勒超声方法测量的LOP相比,这两种方法在临床上都具有可接受的准确性,但这两种自动方法尚未进行直接比较。本研究的目的是调查嵌入式和远端LOP测量方法在临床上是否具有可接受的一致性。使用改良的布兰德和奥特曼分析比较了81名健康个体上下肢LOP测量值对之间的差异。在手术中,由于肢体操作,袖带压力偏离设定点是很常见的。手术止血带系统使用±15 mmHg的压力报警窗口,即如果袖带压力偏离设定点>15 mmHg,就会触发视听警报。因此,如果嵌入式和远端方法在LOP测量中的差异(偏差)±标准误、偏差的95%置信区间以及差异的标准差±标准误均在±15 mmHg以内,这将表明这两种方法在临床上具有可接受的一致性。
使用嵌入式LOP方法测量的LOP比远端LOP方法低-0.81±0.75 mmHg(偏差±标准误)。偏差的95%置信区间为-2.29至0.66 mmHg。差异的标准差±标准误为10.35±0.49 mmHg。这些结果表明,嵌入式和远端LOP测量方法在临床上具有可接受的一致性。
本研究结果表明,嵌入式和远端LOP测量方法在临床上具有可接受的一致性。这些发现支持使用使用两用止血带袖带的嵌入式LOP自动测量方法,以便在临床环境中准确、有效且简单地开具个性化止血带袖带压力处方。