Picciolo G L, Chappelle E W, Thomas R R, McGarry M A
Appl Environ Microbiol. 1977 Dec;34(6):720-4. doi: 10.1128/aem.34.6.720-724.1977.
The performance characteristics of a new photometer that incorporated a reaction system consisting of a movable filtration tape have been studied for use with the firefly luciferase assay for adenosine 5'-triphosphate. Precision, linearity, and sensitivity are given for measuring a graded series of constant-light emitters, concentrations of adenosine 5'-triphosphate, and washed bacterial cells. Precision ranged from 1 to 10% coefficient of variation. The coefficient of correlation was 0.9985 for 7 X 10(8) to 7 X 10(3) bacteria per ml; however, the accuracy decreased at the low levels. The sample processing time was 2 min for a 1-ml sample. This instrument shows potential for many applications in quantitating small numbers of organisms, e.g., pollution, water monitoring, and clinical infection detection.
一种新型光度计的性能特征已得到研究,该光度计采用了由可移动过滤带组成的反应系统,用于三磷酸腺苷的萤火虫荧光素酶测定。给出了测量一系列分级恒定发光体、三磷酸腺苷浓度和洗涤后的细菌细胞时的精密度、线性和灵敏度。变异系数的精密度范围为1%至10%。每毫升7×10⁸至7×10³个细菌时,相关系数为0.9985;然而,在低水平时准确性会降低。对于1毫升样品,样品处理时间为2分钟。该仪器在定量少量生物体方面有许多应用潜力,例如污染、水监测和临床感染检测。