Clinical Technology Service, Neonatal and Paediatric Intensive Care Units, Royal Children's Hospital, Parkville, Victoria, Australia.
Respir Care. 2021 Aug;66(8):1254-1262. doi: 10.4187/respcare.08408. Epub 2021 May 18.
Many pediatric and neonatal ICU patients receive nitric oxide (NO), with some also requiring magnetic resonance imaging (MRI) scans. MRI-compatible NO delivery devices are not always available. We describe and bench test a method of delivering NO during MRI using standard equipment in which a NO delivery device was positioned in the MRI control room with the NO blender component connected to oxygen and set to 80 ppm and delivering flow via 12 m of tubing to a MRI-compatible ventilator, set up inside the MRI scanner magnet room.
For our bench test, the ventilator was set up normally and connected to an infant test lung to simulate several patients of differing weight (ie, 4 kg, 10 kg, 20 kg). The NO blender delivered flows of 2-10 L/min to the ventilator to achieve a range of NO and oxygen concentrations monitored via extended tubing. The measured values were compared to calculated values.
A range of NO concentrations (12-41 ppm) and values (0.67-0.97) were achieved during the bench testing. The additional flow increased delivered peak inspiratory pressure and PEEP by 1-5 cm HO. Calculated values were within acceptable ranges and were used to create a lookup table.
In clinical use, this system can safely generate a range of NO flows of 15-42 ppm with an accompanying range of 0.34-0.98.
许多儿科和新生儿 ICU 患者接受一氧化氮(NO)治疗,其中一些患者还需要磁共振成像(MRI)扫描。并非所有 MRI 兼容的 NO 输送设备都可用。我们描述并在台架上测试了一种在 MRI 期间使用标准设备输送 NO 的方法,其中将 NO 输送设备放置在 MRI 控制室中,将 NO 混合器组件连接到氧气上,并设置为 80ppm,并通过 12m 长的管道输送至 MRI 兼容的呼吸机,该呼吸机设置在 MRI 扫描仪磁体室内。
对于我们的台架测试,呼吸机正常设置,并连接到婴儿测试肺,以模拟不同体重的几个患者(即 4kg、10kg、20kg)。NO 混合器向呼吸机输送 2-10L/min 的流量,以实现通过延长管道监测的一系列 NO 和氧气浓度。测量值与计算值进行了比较。
在台架测试期间,实现了一系列的 NO 浓度(12-41ppm)和 值(0.67-0.97)。额外的流量将输送的峰值吸气压力和 PEEP 增加了 1-5cmHO。计算值在可接受的范围内,并用于创建查找表。
在临床应用中,该系统可以安全地产生 15-42ppm 的一系列 NO 流量,同时伴随 0.34-0.98 的 值。