In our laboratory, the negative rapid group A streptococcal (GAS) antigen assays are backed up by the Solana® GAS Assay by Quidel instead of a Group A streptococcal throat culture. Another FDA cleared RT-PCR assay is the Xpert® Xpress Strep A, which detects Streptococcus pyogenes DNA, and is performed on the Cepheid GeneXpert instrument. Three hundred seventy-five positive and negative specimens were randomly selected from 5489 throat specimens that had been tested by the Solana® GAS Assay during January 2018 and were tested with the Xpress Strep A assay. A throat culture was also set up (sheep blood agar at 35 °C in 5% CO2). All beta-hemolytic streptococci were purified and identified by MALDI-TOF mass spectrometry. Of the 375 samples, 185 were positive by Solana® GAS Assay, and 187 were positive by the Xpress Strep A. The total agreement between the Solana® GAS Assay and the Xpert® Xpress Strep A was 99.5%. The agreement of the Xpert® Xpress Strep A assay with culture was 90.1%. The sensitivity and specificity for Xpress Strep A versus culture were 100% and 83.5%, respectively. The Xpert® Xpress Strep A assay's performance was equivalent to the Solana® GAS Assay, and was highly sensitive. The lower specificity was likely due to the Xpress Strep A assay having higher sensitivity as compared to throat culture.
在我们的实验室中,阴性快速 A 组链球菌(GAS)抗原检测由 Quidel 的 Solana® GAS 检测替代 A 组链球菌咽拭子培养。另一种获得 FDA 批准的实时 RT-PCR 检测是 Xpert® Xpress Strep A,它可检测化脓性链球菌 DNA,在 Cepheid GeneXpert 仪器上进行。从 2018 年 1 月使用 Solana® GAS 检测的 5489 例咽拭子标本中随机选择了 375 例阳性和阴性标本,并用 Xpress Strep A 检测进行检测。还建立了咽拭子培养(羊血琼脂在 35°C 和 5% CO2 下)。所有乙型溶血性链球菌均通过 MALDI-TOF 质谱进行纯化和鉴定。在 375 个样本中,Solana® GAS 检测有 185 个阳性,Xpress Strep A 有 187 个阳性。Solana® GAS 检测与 Xpert® Xpress Strep A 的总符合率为 99.5%。Xpert® Xpress Strep A 检测与培养的符合率为 90.1%。Xpress Strep A 对培养的敏感性和特异性分别为 100%和 83.5%。Xpert® Xpress Strep A 检测的性能与 Solana® GAS 检测相当,且具有很高的敏感性。特异性较低可能是由于 Xpress Strep A 检测的敏感性高于咽拭子培养。