Suppr超能文献

新型 DXA 系统 Shimazu SONIALVISION G4 与 GE-Lunar PRODIGY 的比较。

Comparison of novel DXA system, Shimazu SONIALVISION G4, with GE-Lunar PRODIGY.

机构信息

Department of Orthopedic Surgery, Anjo Kosei Hospital, 28 Higashihirokute, Anjo City, Aichi, Japan.

Department of Radiological Technology, Anjo Kosei Hospital, 28 Higashihirokute, Anjo City, Aichi, Japan.

出版信息

J Bone Miner Metab. 2021 Nov;39(6):1041-1047. doi: 10.1007/s00774-021-01241-2. Epub 2021 Jun 11.

Abstract

INTRODUCTION

SONIALVISION G4 (Shimadzu, Kyoto, Japan) is a fluoroscopic imaging system capable of dual-energy X-ray absorptiometry (DXA) using the additional application. However, bone mineral density (BMD) with this system is calibrated only by a phantom, and the necessity to confirm the reliability of BMD measurement in human subjects has been pointed out. This study aimed to evaluate the clinical reliability of BMD measurement using SONIALVISION G4 by comparing it with PRODIGY (GE Healthcare, Madison, WI, USA), a reference system in developing the DXA application.

MATERIALS AND METHODS

BMD measurements using these two DXA systems were performed on the same day at the lumbar spine and bilateral proximal femur of 130 subjects aged 22-86 years. The agreement and correlation between the measurements were determined. Thirty out of the 130 subjects were scanned twice to determine the precision of the SONIALVISION G4 system.

RESULTS

BMD was highly correlated between the two DXA systems both at the lumbar spine and proximal femur, with correlation coefficients ranging from 0.963 (left total hip) to 0.989 (left femoral neck). Differences in BMD measured by the two systems were less than 3.0%, but a small proportional bias was observed in the Bland-Altman analysis. Coefficients of variation in BMD measurement were 1.77% in the lumbar spine and 1.41-2.76% in the proximal femur.

CONCLUSIONS

Although BMD values by SONIALVISION G4 were close to those by PRODIGY, follow-up studies should be performed using the same device because of the small difference between devices.

摘要

简介

SONIALVISION G4(日本岛津制作所)是一款能够通过附加应用程序进行双能 X 射线吸收法(DXA)的透视成像系统。然而,该系统的骨矿物质密度(BMD)仅通过体模进行校准,因此已经指出需要确认其在人体受试者中测量 BMD 的可靠性。本研究旨在通过与 PRODIGY(美国通用电气医疗集团,麦迪逊,威斯康星州)进行比较,来评估 SONIALVISION G4 系统进行 BMD 测量的临床可靠性,PRODIGY 是开发 DXA 应用程序的参考系统。

材料和方法

在 130 名年龄在 22-86 岁的受试者的同一天,使用这两种 DXA 系统对腰椎和双侧股骨近端进行 BMD 测量。确定了两种测量方法之间的一致性和相关性。从 130 名受试者中选择 30 名进行两次扫描,以确定 SONIALVISION G4 系统的精度。

结果

两种 DXA 系统在腰椎和股骨近端的 BMD 均高度相关,相关系数范围从 0.963(左总髋部)到 0.989(左股骨颈)。两种系统测量的 BMD 差异小于 3.0%,但在 Bland-Altman 分析中观察到小的比例偏差。腰椎 BMD 测量的变异系数为 1.77%,股骨近端为 1.41-2.76%。

结论

尽管 SONIALVISION G4 的 BMD 值与 PRODIGY 接近,但由于设备之间存在微小差异,应使用相同的设备进行随访研究。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验