Porous-coated femoral components in a canine model for revision arthroplasty.

作者信息

Roberson J R, Spector M, Baggett M A, Kita K

机构信息

Department of Orthopaedics, Emory University School of Medicine, Atlanta, Georgia 30322.

出版信息

J Bone Joint Surg Am. 1988 Sep;70(8):1201-8.

PMID:3417705
Abstract

UNLABELLED

Six dogs had a total hip arthroplasty during which the femoral component was coated with methacrylate and inserted in the femoral canal, after the canal had been reamed to a larger diameter than that of the femoral stem (including the coating). Thus, the implant was loose and motion was present between it and the femur in each animal. Five dogs had a revision to a prosthesis with a porous polysulfone-coated stem. This prosthesis was not cemented in place. One dog was killed before the revision for the purpose of histological examination. Although bone was present in the porous surface of all five stems at the revision, the amount was scant in three.

CLINICAL RELEVANCE

A model is described for the study of cemented hip-replacement prostheses that have failed. Although variable amounts of ingrowth of bone were observed after the revision to the porous-coated prostheses without bone cement, further study is needed to determine whether adequate ingrowth of bone occurs in this model.

摘要

未标记

六只狗接受了全髋关节置换术,术中股骨部件用甲基丙烯酸酯涂层,并在股骨骨髓腔扩孔至比股骨干(包括涂层)直径更大后插入骨髓腔。因此,每只动物的植入物都很松动,其与股骨之间存在活动。五只狗翻修为使用多孔聚砜涂层柄的假体。该假体未用骨水泥固定到位。一只狗在翻修前被处死用于组织学检查。尽管翻修时所有五个柄的多孔表面都有骨长入,但其中三个的骨量很少。

临床相关性

描述了一种用于研究失败的骨水泥型髋关节置换假体的模型。尽管在翻修至无骨水泥的多孔涂层假体后观察到不同程度的骨长入,但仍需要进一步研究以确定该模型中是否会发生足够的骨长入。