Hallock G G
Division of Plastic Surgery, Allentown Affiliated Hospital, Pa.
Plast Reconstr Surg. 1988 Oct;82(4):666-8. doi: 10.1097/00006534-198810000-00018.
An ex vivo study was devised to ascertain the maximum number of needle punctures possible prior to onset of leakage from remote-type tissue expansion reservoirs. At least 10 standard valves obtained from all major manufacturers were analyzed using various needle gauges. An 18-gauge beveled needle was determined to be the maximum size that could be safely employed for all vendors within a usual course for tissue expansion. No significant difference was found in comparison of Dacron-reinforced versus plain silicone membrane-type valves.