Respiratory motion may compromise the dose delivery accuracy in liver stereotactic body radiation therapy (SBRT). Motion management can improve treatment delivery. However, external surrogate signal may be unstable and inaccurate. This study reports the first case of liver SBRT based on internal electromagnetic motion monitoring (Calypso, Varian Medical Systems, USA) in China.
The patient with a primary liver cancer was treated with respiratory-gated SBRT guided by three implanted electromagnetic transponders. The treatment was carried out in breath-hold end-exhale with beam-on when the centroid of the three transponders drifted within 5 mm (left-right (LR), anterior-posterior (AP) and cranio-caudal (CC) directions) from the planned position. The motion monitoring treatments were delivered in breath-hold end-exhale mode with the energy of 6 MV in FFF mode with 1200 monitor units (MU) per minute. For each fraction, QA results, intertransponder distances, geometric checks as well as tumor motion logs were explicitly recorded.
Comparing with the plan data, distance variances between each two transponders were - 0.56 ± 0.32 mm, 0.17 ± 0.33 mm and - 0.82 ± 0.68 mm. Geometric residual, the pitch, roll and yaw angles were 0.48 ± 0.21 mm (threshold 2.0 mm), 2.17° ± 1.85° (threshold 10°), - 2.42° ± 1.51° (threshold 10°) and 1.67° ± 1.07° (threshold 10°), respectively. The delivery time of the five fields were 13.8 s, 13.1 s, 11.2 s, 11.6 s, and 11.6 s with the average value of 12.3 ± 1.1 s. Treatment duration of each fraction ranged from 6.2 to 21.4 min, with the average value of 11.3 ± 5.0 min.
The first case of liver SBRT patient of China based on internal electromagnetic motion monitoring was performed. The system had a high tracking accuracy, and it did not delay the treatment time. In addition, the patient did not show any severe side effects except for grade I myelotoxicity. The internal electromagnetic motion monitoring system provides a real-time and direct way to track liver tumor targets.
呼吸运动会影响肝脏立体定向体部放射治疗(SBRT)的剂量分布精度。运动管理可以提高治疗效果。但是,外部的替代信号可能不稳定和不准确。本研究报道了首例基于内部电磁运动监测(美国瓦里安医疗系统的 Calypso)的中国肝脏 SBRT 病例。
采用三维植入电磁传感器引导原发性肝癌患者接受呼吸门控 SBRT 治疗。在治疗中,当三个传感器的质心在计划位置左右(LR)、前后(AP)和头尾(CC)方向上漂移超过 5mm 时,使用束流 ON 的屏气末呼气技术进行治疗。采用能量为 6MV 的 FFF 模式,每个 MU 每分钟 1200MU,以屏气末呼气模式进行运动监测治疗。对每个分次,明确记录 QA 结果、传感器之间的距离、几何检查和肿瘤运动日志。
与计划数据相比,两个传感器之间的距离偏差分别为-0.56±0.32mm、0.17±0.33mm 和-0.82±0.68mm。几何残余、俯仰、滚转和偏航角分别为 0.48±0.21mm(阈值 2.0mm)、2.17°±1.85°(阈值 10°)、-2.42°±1.51°(阈值 10°)和 1.67°±1.07°(阈值 10°)。五个射野的传输时间分别为 13.8s、13.1s、11.2s、11.6s 和 11.6s,平均为 12.3±1.1s。每个分次的治疗时间为 6.2-21.4min,平均为 11.3±5.0min。
首例中国基于内部电磁运动监测的肝脏 SBRT 患者完成治疗。该系统具有较高的跟踪精度,且不会延迟治疗时间。此外,患者除了 I 级骨髓毒性外,没有任何严重的副作用。内部电磁运动监测系统为实时、直接地跟踪肝脏肿瘤靶区提供了一种方法。