J Electrocardiol. 2021 Jul-Aug;67:148-157. doi: 10.1016/j.jelectrocard.2021.06.009. Epub 2021 Jul 2.
INTRODUCTION: Photoplethysmography (PPG) in wearable sensors potentially plays an important role in accessible heart rhythm monitoring. We investigated the accuracy of a state-of-the-art bracelet (Corsano 287) for heartbeat detection in cardiac patients and evaluated the efficacy of a signal qualifier in identifying medically useful signals. METHODS: Patients from an outpatient cardiology clinic underwent a simultaneous resting ECG and PPG recording, which we compared to determine accuracy of the PPG sensor for detecting heartbeats within 100 and 50 ms of the ECG-detected heart beats and correlation and Limits of Agreement for heartrate (HR) and RR-intervals. We defined subgroups for skin type, hair density, age, BMI and gender and applied a previously described signal qualifier. RESULTS: In 180 patients 7914 ECG-, and 7880 (99%) PPG-heartbeats were recorded. The PPG-accuracy within 100 ms was 94.6% (95% CI 94.1-95.1) and 89.2% (95% CI 88.5-89.9) within 50 ms. Correlation was high for HR (R = 0.991 (95% CI 0.988-0.993), n = 180) and RR-intervals (R = 0.891 (95% CI 0.886-0.895), n = 7880). The 95% Limits of Agreement (LoA) were -3.89 to 3.77 (mean bias 0.06) beats per minute for HR and -173 to 171 (mean bias -1) for RR-intervals. Results were comparable across all subgroups. The signal qualifier led to a higher accuracy in a 100 ms range (98.2% (95% CI 97.9-98.5)) (n = 143). CONCLUSION: We showed that the Corsano 287 Bracelet with PPG-technology can determine HR and RR-intervals with high accuracy in cardiovascular at-risk patient population among different subgroups, especially with a signal quality indicator.
简介:可穿戴传感器中的光电容积脉搏波描记法(PPG)在可访问的心率监测中可能发挥着重要作用。我们研究了一种最先进的手镯(Corsano 287)在心脏病人群中检测心跳的准确性,并评估了信号质量指标在识别有医疗价值信号方面的效果。
方法:来自门诊心内科的患者接受了同步静息心电图和 PPG 记录,我们将其进行比较,以确定 PPG 传感器在检测心电图检测到的心跳 100 毫秒和 50 毫秒内的准确性,并评估心率(HR)和 RR 间隔的相关性和界限协议。我们根据皮肤类型、毛发密度、年龄、BMI 和性别定义了亚组,并应用了之前描述的信号质量指标。
结果:在 180 名患者中,记录了 7914 次心电图和 7880 次(99%)PPG 心跳。PPG 在 100 毫秒内的准确率为 94.6%(95%置信区间 94.1-95.1),在 50 毫秒内的准确率为 89.2%(95%置信区间 88.5-89.9)。HR(R = 0.991(95%置信区间 0.988-0.993),n = 180)和 RR 间隔(R = 0.891(95%置信区间 0.886-0.895),n = 7880)的相关性很高。HR 的 95%界限协议(LoA)为-3.89 至 3.77(平均偏差 0.06)次/分钟,RR 间隔的 LoA 为-173 至 171(平均偏差-1)次/分钟。结果在所有亚组中均具有可比性。信号质量指标在 100 毫秒范围内可提高准确率(98.2%(95%置信区间 97.9-98.5))(n = 143)。
结论:我们表明,Corsano 287 腕带与 PPG 技术相结合,可在不同亚组的心血管高危患者人群中以高精度确定 HR 和 RR 间隔,特别是使用信号质量指标时。