连续光电容积脉搏波监测 1 分钟平均心率在心房颤动中的准确性评估。

Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation.

机构信息

Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht, P. Debyelaan 25, 6229 HX Maastricht, The Netherlands.

Department of Biomedical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Nørregade 10, 1165 Copenhagen, Denmark.

出版信息

Europace. 2023 Mar 30;25(3):835-844. doi: 10.1093/europace/euad011.

Abstract

AIMS

Although mobile health tools using photoplethysmography (PPG) technology have been validated for the detection of atrial fibrillation (AF), their utility for heart rate assessment during AF remains unclear. Therefore, we aimed to evaluate the accuracy of continuous PPG-based 1 min mean heart rate assessment during AF.

METHODS AND RESULTS

Persistent AF patients were provided with Holter electrocardiography (ECG) (for ≥24 h) simultaneously with a PPG-equipped smartwatch. Both the PPG-based smartwatch and Holter ECG automatically and continuously monitored patients' heart rate/rhythm. ECG and PPG recordings were synchronized and divided into 1 min segments, from which a PPG-based and an ECG-based average heart rate estimation were extracted. In total, 47 661 simultaneous ECG and PPG 1 min heart rate segments were analysed in 50 patients (34% women, age 73 ± 8 years). The agreement between ECG-determined and PPG-determined 1 min mean heart rate was high [root mean squared error (RMSE): 4.7 bpm]. The 1 min mean heart rate estimated using PPG was accurate within ±10% in 93.7% of the corresponding ECG-derived 1 min mean heart rate segments. PPG-based 1 min mean heart rate estimation was more often accurate during night-time (97%) than day-time (91%, P < 0.001) and during low levels (96%) compared to high levels of motion (92%, P < 0.001). A neural network with a 10 min history of the recording did not further improve the PPG-based 1 min mean heart rate assessment [RMSE: 4.4 (95% confidence interval: 3.5-5.2 bpm)]. Only chronic heart failure was associated with a lower agreement between ECG-derived and PPG-derived 1 min mean heart rates (P = 0.040).

CONCLUSION

During persistent AF, continuous PPG-based 1 min mean heart rate assessment is feasible in 60% of the analysed period and shows high accuracy compared with Holter ECG for heart rates <110 bpm.

摘要

目的

虽然使用光电容积脉搏波(PPG)技术的移动健康工具已被验证可用于检测心房颤动(AF),但其在 AF 期间评估心率的效用尚不清楚。因此,我们旨在评估在 AF 期间连续使用 PPG 进行 1 分钟平均心率评估的准确性。

方法和结果

持续性 AF 患者同时佩戴配备 PPG 的智能手表进行 24 小时以上的动态心电图(ECG)监测。PPG 智能手表和动态心电图自动连续监测患者的心率/节律。将心电图和 PPG 记录同步并分为 1 分钟段,从中提取基于 PPG 和基于 ECG 的平均心率估计值。共分析了 50 名患者(34%为女性,年龄 73±8 岁)的 47661 个同时进行的 ECG 和 PPG 1 分钟心率段。ECG 确定的 1 分钟平均心率与 PPG 确定的 1 分钟平均心率之间具有高度一致性[均方根误差(RMSE):4.7 bpm]。在 93.7%的相应 ECG 衍生的 1 分钟平均心率段中,PPG 估计的 1 分钟平均心率在±10%范围内准确。在夜间(97%)比白天(91%,P<0.001)和低运动水平(96%)比高运动水平(92%,P<0.001)时,PPG 基于 1 分钟平均心率的估计更准确。具有 10 分钟记录历史的神经网络并没有进一步提高 PPG 基于 1 分钟平均心率的评估[RMSE:4.4(95%置信区间:3.5-5.2 bpm)]。只有慢性心力衰竭与 ECG 衍生的和 PPG 衍生的 1 分钟平均心率之间的一致性较低有关(P=0.040)。

结论

在持续性 AF 中,在分析期间的 60%时段内可以连续进行基于 PPG 的 1 分钟平均心率评估,并且与 Holter ECG 相比,对于<110 bpm 的心率具有很高的准确性。