多管体模的使用:一种全局校正专用分子磁共振乳腺线圈中固有T1弛豫时间不均匀性的方法。

Use of Multiple-Tube Phantom: A Method to Globally Correct Native T1 Relaxation Time Inhomogeneity in Dedicated Molecular Magnetic Resonance Breast Coil.

作者信息

Negi Pradeep Singh, Mehta Shashi Bhushan, Jena Amarnath

机构信息

Department of Molecular Imaging and Nuclear Medicine, PET SUITE, Indraprastha Apollo Hospitals, New Delhi, India.

Department of Physics, Vivekananda Global University, Jaipur, Rajasthan, India.

出版信息

J Med Phys. 2021 Jan-Mar;46(1):41-46. doi: 10.4103/jmp.JMP_2_20. Epub 2021 May 5.

Abstract

BACKGROUND

Native T1 relaxation time (T1) presents an important prerequisite to reliably quantify pharmacokinetic parameter like K (volume transfer constant). Native T1 value can be varied because of the inhomogeneity in the breast coil, thus influencing the K measurement.

PURPOSE

The current study aims to design and use a phantom with multiple tubes for both breast cuffs to assess native T1 inhomogeneity across the dedicated molecular magnetic resonance (mMR) breast coil and adopt corrective method to spatially normalize T1 values to improve homogeneity.

MATERIALS AND METHODS

Two phantoms with multiple tubes (19 tubes) specially designed and filled with contrast medium with known T1 value were placed in each mMR breast coil cuff. Native T1 at various spatial locations was calculated applying dual flip angle sequence. Correction factors were derived at various spatial locations as a function of deviation of the native T1 value from phantom and applied to correct the native T1 relaxation time.

RESULTS

A statistically significant difference between native T1 values of the right and left anterior ( = 0.0095), middle ( = 0.0081), and posterior ( = 0.0004) parts of the breast coil. No significant difference was seen in the corrected T1 values between anterior ( = 0.402), middle ( = 0.305), and posterior ( = 0.349) aspects of both sides of the breast coil.

CONCLUSION

Inhomogeneity in the native T1 value exists in dedicated mMR breast coil, and significant improvement can be achieved using specially designed external phantom with multiple tubes.

摘要

背景

固有T1弛豫时间(T1)是可靠量化诸如K(容积转移常数)等药代动力学参数的重要前提。由于乳腺线圈的不均匀性,固有T1值可能会发生变化,从而影响K值的测量。

目的

本研究旨在设计并使用一种带有多根管子的体模,用于双侧乳腺线圈,以评估专用分子磁共振(mMR)乳腺线圈上的固有T1不均匀性,并采用校正方法对T1值进行空间归一化,以提高均匀性。

材料与方法

将两个专门设计的带有多根管子(19根管子)并填充有已知T1值造影剂的体模放置在每个mMR乳腺线圈套囊中。应用双翻转角序列计算不同空间位置的固有T1。根据固有T1值与体模的偏差,在不同空间位置得出校正因子,并应用于校正固有T1弛豫时间。

结果

乳腺线圈左右前侧(P = 0.0095)、中间(P = 0.0081)和后侧(P = 0.0004)的固有T1值之间存在统计学显著差异。乳腺线圈两侧前侧(P = 0.402)、中间(P = 0.305)和后侧(P = 0.349)的校正T1值之间无显著差异。

结论

专用mMR乳腺线圈中存在固有T1值不均匀性,使用专门设计的带有多根管子的外部体模可实现显著改善。